FDA Enforcement Class II Terminated

Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571

Recall: Z-2297-2020 · Reported June 17, 2020

Enforcement

Recall Number
Z-2297-2020
Event ID
85600
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Elekta Limited
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 17, 2020
Initiation Date
May 21, 2020
Classification Date
June 8, 2020
Termination Date
August 19, 2024
Address
Linac House, Fleming Way, Crawley, N/A, N/A, United Kingdom

Description

Elekta Unity, Philips Marlin Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571

Reason

In TSM (Treatment Session Manager) Motion Monitoring workflows, under certain conditions there can be a mismatch between the contour data overlay with respect to the motion monitoring images of the monitored structure.

Code Info

Software versions: R5.3.30, R5.3.31, R6.0.5331, R5.7.0, R5.7.1, R6.1.571. UDI Number 05060191071321. Anatomy (Protocol Name): Abdomen (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Fast Sag Cor RealTime, btFFE Sag RealTime); HeadNeck L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); HeadNeck M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Pelvis L (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Pelvis M (bFFE Cor RealTime, bFFE Cor Sag RealTime, bFFE Cor Sag Tra RealTime, bFFE Sag RealTime, bFFE Sag Tra RealTime); Thorax (btFFE Cor RealTime, btFFE Cor Sag RealTime, btFFE Cor Sag Tra RealTime, btFFE Sag RealTime, btFFE Sag RealTime)

Distribution

US distribution to IA, NY, PA, TX, WI. International distribution to Australia, Canada, China, Denmark, Germany, Hong Kong, Italy, Netherlands, Sweden, Switzerland, Turkey, United Kingdom.

Quantity

26 units