FDA Enforcement Class II Ongoing

Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo, and Atopic Dermatitis (Eczema) under the direction of a physician, Model Numbers 210NDT4 and 210NCL4

Recall: Z-2295-2026 · Reported June 10, 2026

Enforcement

Recall Number
Z-2295-2026
Event ID
98817
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Daavlin Distributing Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 10, 2026
Initiation Date
December 10, 2025
Classification Date
June 2, 2026
Address
205 W Bement St, Bryan, OH, 43506-1264, United States

Description

Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo, and Atopic Dermatitis (Eczema) under the direction of a physician, Model Numbers 210NDT4 and 210NCL4

Reason

Some of the Calibration/Output Certificates had the outputs reversed.

Code Info

Serial Numbers: 2-03409NB4DT 2-03560NB4DT 2-03798NB4DT 2-03472NB4DT 2-03315NB4DT 2-03726NB4DT 2-03396NB4DT 2-03320NB4DT 2-03447NB4DT 2-03564NB4DT 2-03516NB4DT 2-03467NB4DT 2-03292NB4DT 2-03371NB4DT 2-03433NB4DT 2-03446NB4DT 2-03343NB4DT 2-03400NB4DT 2-03342NB4DT 2-03547NB4DT 2-03449NB4DT 2-03685NB4DT 2-03574NB4DT 2-03392NB4DT 2-03345NB4DT 2-03431NB4DT 2-03608NB4DT 2-03530NB4DT 2-03501NB4DT 2-03362NB4DT 2-03603NB4DT 2-03713NB4DT 2-03508NB4DT 2-03341NB4DT 2-03418NB4DT 2-03721NB4DT 2-03662NB4DT 2-03679NB4DT 2-03591NB4DT 2-03319NB4DT 2-03411NB4DT 2-03505NB4DT 2-03335NB4DT 2-03585NB4DT 2-03556NB4DT 2-03355NB4DT 2-03664NB4DT 2-03515NB4DT 2-03424NB4DT 2-03470NB4DT 2-03386NB4DT 2-03455NB4DT 2-03401NB4DT 2-03391NB4DT 2-03578NB4DT 2-03388NB4DT 2-03632NB4DT 2-03615NB4DT 2-03434NB4DT 2-03454NB4DT 2-03387NB4DT 2-03390NB4DT 2-03569NB4DT 2-03541NB4DT 2-03509NB4DT 2-03789NB4DT 2-03804NB4DT 2-03439NB4DT 2-03367NB4DT 2-03442NB4DT 2-03750NB4DT 2-03349NB4DT 2-03510NB4DT 2-03538NB4DT 2-03559NB4DT 2-03539NB4DT 2-03491NB4DT 2-03502NB4DT 2-03629NB4DT 2-03383NB4DT 2-03496NB4DT 2-03619NB4DT 2-03511NB4DT 2-03670NB4DT 2-03565NB4DT 2-03581NB4DT 2-03479NB4DT 2-03348NB4DT 2-03534NB4DT 2-03484NB4DT 2-03802NB4DT 2-03504NB4DT 2-03457NB4DT 2-03417NB4DT 2-03482NB4DT 2-03318NB4DT 2-03503NB4DT 2-03518NB4DT 2-03350NB4DT 2-03686NB4DT 2-03317NB4DT 2-03687NB4DT 2-03507NB4DT 2-03663NB4DT 2-03658NB4DT 2-03330NB4DT 2-03489NB4DT 2-03451NB4DT 2-03486NB4DT 2-03465NB4DT 2-03346NB4DT 2-03598NB4DT 2-03377NB4DT 2-03414NB4DT 2-03422NB4DT 2-03531NB4DT 2-03329NB4DT 2-03579NB4DT 2-03817NB4DT 2-03613NB4DT 2-03322NB4DT 2-03570NB4DT 2-03397NB4DT 2-03326NB4DT 2-03347NB4DT 2-03469NB4DT 2-03483NB4DT 2-03359NB4DT 2-03344NB4DT 2-03436NB4DT 2-03711NB4DT 2-03437NB4DT 2-03459NB4DT 2-03384NB4DT 2-03389NB4DT 2-03592NB4DT 2-03432NB4DT 2-03823NB4DT 2-03616NB4DT 2-03498NB4DT 2-03512NB4DT 2-03809NB4DT 2-03647NB4DT 2-03606NB4DT 2-03722NB4DT 2-03818NB4DT 2-03589NB4DT 2-03356NB4DT 2-03583NB4DT 2-03799NB4DT 2-03487NB4DT 2-03398NB4DT 2-03775NB4DT 2-03427NB4DT 2-03533NB4DT 2-03676NB4DT 2-03468NB4DT 2-03684NB4DT 2-03413NB4DT 2-03716NB4DT 2-03374NB4DT 2-03430NB4DT 2-03473NB4DT 2-03575NB4DT 2-03339NB4DT 2-03666NB4DT 2-03323NB4DT 2-03471NB4DT 2-03334NB4DT 2-03310NB4DT 2-03631NB4DT 2-03405NB4DT 2-03720NB4DT 2-03372NB4DT 2-03463NB4DT 2-03353NB4DT 2-03563NB4DT 2-03419NB4DT 2-03725NB4DT 2-03621NB4DT 2-03336NB4DT 2-03312NB4DT 2-03806NB4DT 2-03572NB4DT 2-03597NB4DT 2-03370NB4DT 2-03435NB4DT 2-03354NB4DT 2-03555NB4DT 2-03440NB4DT 2-03376NB4DT 2-03610NB4DT 2-03653NB4DT 2-03749NB4DT 2-03380NB4DT 2-03558NB4DT 2-03462NB4DT 2-03313NB4DT 2-03517NB4DT 2-03790NB4DT 2-03513NB4DT 2-03602NB4DT 2-03683NB4DT 2-03561NB4DT 2-03514NB4DT 2-03604NB4DT 2-03445NB4DT 2-03423NB4DT 2-03338NB4DT 2-03375NB4DT 2-03361NB4DT 2-03543NB4DT 2-03492NB4DT 2-03801NB4DT 2-03497NB4DT 2-03601NB4DT 2-03824NB4DT 2-03477NB4DT 2-03545NB4DT 2-03813NB4DT 2-03557NB4DT 2-03426NB4DT 2-03428NB4DT 2-03394NB4DT 2-03645NB4DT 2-03842NB4DT 2-03528NB4DT 2-03321NB4DT 2-03478NB4DT 2-03438NB4DT 2-03369NB4DT 2-03661NB4DT 2-03682NB4DT 2-03412NB4DT 2-03351NB4DT 2-03673NB4DT 2-03453NB4DT 2-03605NB4DT 2-03461NB4DT 2-03495NB4DT 2-03582NB4DT 2-03540NB4DT 2-03594NB4DT 2-03612NB4DT 2-03810NB4DT 2-03364NB4DT 2-03609NB4DT 2-03672NB4DT 2-03529NB4DT 2-03379NB4DT 2-03554NB4DT 2-03490NB4DT 2-03407NB4DT 2-03586NB4DT 2-03600NB4DT 2-03618NB4DT 2-03421NB4DT 2-03659NB4DT 2-03671NB4DT 2-03448NB4DT 2-03607NB4DT 2-03402NB4DT 2-03485NB4DT 2-03480NB4DT 2-03420NB4DT 2-03328NB4DT 2-03408NB4DT 2-03481NB4DT 2-03625NB4DT 2-03331NB4DT 2-03614NB4DT 2-03646NB4DT 2-03668NB4DT 2-03800NB4DT 2-03499NB4DT 2-03753NB4DT 2-03520NB4DT 2-03464NB4DT 2-03316NB4DT 2-03580NB4DT 2-03675NB4DT 2-03665NB4DT 2-03546NB4DT 2-03590NB4DT 2-03544NB4DT 2-03415NB4DT 2-03719NB4DT 2-03456NB4DT 2-03532NB4DT 2-03385NB4DT 2-03452NB4DT 2-03593NB4DT 2-03366NB4DT 2-03759NB4DT 2-03506NB4DT 2-03425NB4DT 2-03660NB4DT 2-03360NB4DT 2-03519NB4DT 2-03688NB4DT 2-03537NB4DT 2-03429NB4DT 2-03406NB4DT 2-03340NB4DT 2-03311NB4DT 2-03562NB4DT 2-03474NB4DT 2-03655NB4DT 2-03466NB4DT 2-03595NB4DT 2-03476NB4DT 2-03403NB4DT 2-03752NB4DT 2-03332NB4DT 2-03314NB4DT 2-03309NB4DT 2-03678NB4DT 2-03410NB4DT 2-03460NB4DT 2-03717NB4DT 2-03521NB4DT 2-03674NB4DT 2-03363NB4DT 2-03667NB4DT 2-03475NB4DT 2-03724NB4DT 2-03493NB4DT 2-03617NB4DT 2-03680NB4DT 2-03494NB4DT 2-03450NB4DT 2-03754NB4DT 2-03378NB4DT 2-03712NB4DT 2-03840NB4DT 2-03441NB4DT 2-03416NB4DT 2-03393NB4DT 2-03577NB4DT 2-03751NB4DT 2-03681NB4DT 2-03368NB4DT 2-03333NB4DT 2-03640NB4DT 2-03373NB4DT 2-03358NB4DT 2-03500NB4DT 2-03596NB4DT 2-03811NB4DT 2-03308NB4DT 2-03669NB4DT 2-03337NB4DT 2-03381NB4DT 2-03352NB4DT 2-03488NB4DT 2-03382NB4DT 2-03443NB4DT 2-03357NB4DT 2-03444NB4DT 2-03542NB4DT 2-03599NB4DT 2-03404NB4DT 2-03395NB4DT 2-03365NB4DT 2-03399NB4DT 2-03576NB4DT 2-03458NB4DT 2-03611NB4DT

Distribution

US Nationwide distribution.

Quantity

368 units