Custom Tubing Kit: REF: K04-00417, K04-00963, K04-01029
Enforcement
- Recall Number
- Z-2294-2025
- Event ID
- 97151
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Merit Medical Systems, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 20, 2025
- Initiation Date
- June 5, 2025
- Classification Date
- August 12, 2025
- Address
- 1600 W Merit Pkwy, N/A, South Jordan, UT, 84095-2416, United States
Description
Custom Tubing Kit: REF: K04-00417, K04-00963, K04-01029
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
REF/UDI-DI/Lot(Expiration): K04-00417/00884450046630/I3114667(12/10/2027), I3147550(1/28/2028); K04-00963/00884450046760/H3117053(12/14/2027), H3138590(1/20/2028), H3154108(2/10/2028), H3154786(2/10/2028), H3162244(2/19/2028), H3166789(3/2/2028), H3173526(3/23/2028); K04-01029/00884450148938/H3117059(12/17/2027), H3154112(2/10/2028), H3159017(2/19/2028), H3166777(3/9/2028), H3181994(4/1/2028);
US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.