FDA Enforcement Class II Terminated

Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories. Scanner Label/Labeling: JADAK JDK-2330 REV B 170512-017

Recall: Z-2292-2019 · Reported August 28, 2019

Enforcement

Recall Number
Z-2292-2019
Event ID
81420
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Luminex Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 28, 2019
Initiation Date
June 29, 2018
Classification Date
August 19, 2019
Termination Date
April 28, 2021
Address
12212 Technology Blvd, N/A, Austin, TX, 78727-6101, United States

Description

Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical laboratories. Scanner Label/Labeling: JADAK JDK-2330 REV B 170512-017

Reason

An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order.

Code Info

ARIES System UDI: 00840487101537 ARIES M1 System UDI: 00840487100080 ARIES-M12V1-IVD, ARIES System ARIES-M6V1-IVD, ARIES M1 System CN-0321-01, ARIES System Handheld Barcode Scanner All scanners manufactured after April 2017 and including Part Number JDK-2330 Rev B.

Distribution

Worldwide distribution. US nationwide, Canada, Ireland, Netherlands, Singapore, Thailand, and United Kingdom

Quantity