FDA Enforcement Class II Ongoing

KNEE ARTHROSCOPY PACK Catalog Number: MEKA12R

Recall: Z-2291-2024 · Reported July 17, 2024

Enforcement

Recall Number
Z-2291-2024
Event ID
94765
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
American Contract Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
July 17, 2024
Initiation Date
May 15, 2024
Classification Date
July 5, 2024
Address
85 Shaffer Park Dr, N/A, Tiffin, OH, 44883-9290, United States

Description

KNEE ARTHROSCOPY PACK Catalog Number: MEKA12R

Reason

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

Code Info

UDI-DI: 00191072174475 LOT#"s: 659231 978241

Distribution

US Nationwide distribution.

Quantity

84 units