FDA Enforcement
Class II
Terminated
Arrow PICC with Chlorag+ard Technology A PICC permits venous access to the central circulation through a peripheral vein. The intended use of the VP'S Stylet and Console (VPS System) is to quickly and accurately guide market available central catheters to the desired location which is the lower third of the SVC or at the cavo-atrial junction
Recall: Z-2291-2017
·
Reported June 14, 2017
Enforcement
- Recall Number
- Z-2291-2017
- Event ID
- 76937
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Arrow International Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 14, 2017
- Initiation Date
- March 20, 2017
- Classification Date
- June 6, 2017
- Termination Date
- December 18, 2018
- Address
- 2400 Bernville Rd, N/A, Reading, PA, 19605-9607, United States
Description
Arrow PICC with Chlorag+ard Technology A PICC permits venous access to the central circulation through a peripheral vein. The intended use of the VP'S Stylet and Console (VPS System) is to quickly and accurately guide market available central catheters to the desired location which is the lower third of the SVC or at the cavo-atrial junction
Reason
Arrow International initiated the recall due to reports of safety cover disengagement and needlestick injury for the BD Eclipse Needle that are packaged with certain Arrow products.
Code Info
Material number CDA-45063-HPK1A
Distribution
Nationwide Distribution
Quantity
unknown-firm reports total devices distributed 43,394