FDA Enforcement Class II Terminated

HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 115V & Model 610HT, 230V Product Usage: Clean and thermally disinfect medical devices using full immersion pasteurization

Recall: Z-2288-2018 · Reported July 4, 2018

Enforcement

Recall Number
Z-2288-2018
Event ID
80295
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cenorin, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
July 4, 2018
Initiation Date
April 9, 2018
Classification Date
June 26, 2018
Termination Date
September 13, 2019
Address
6324 S 199th Pl, Suite 107, Kent, WA, 98032, United States

Description

HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 115V & Model 610HT, 230V Product Usage: Clean and thermally disinfect medical devices using full immersion pasteurization

Reason

There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.

Code Info

Serial Numbers: 611503, 611504, 611505, 611506, 611601, 611602, 611603, 611604, 611605, 611606, 611607, 611608, 611609 & 611610

Distribution

Worldwide Distribution - US Nationwide in the states of CA, FL, IN, NC, NE, UT & WV; and countries of Malaysia and Taiwan

Quantity

14 units