PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilization case is to organize, store and protect its contents from damage consistent with typical hospital or surgical center circulation (i.e. between central services and the operating room) and during transport between a Zimmer distributorship and a health care facility by Zimmer distributorship personnel.
Enforcement
- Recall Number
- Z-2288-2014
- Event ID
- 68653
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 27, 2014
- Initiation Date
- July 1, 2014
- Classification Date
- August 20, 2014
- Termination Date
- June 9, 2015
- Address
- 1800 W Center St, Warsaw, IN, 46580-2304, United States
Description
PERSONA (TM) The Personalized Knee System Tibial General Instrument Tray, non-sterile The intended use of the sterilization case is to organize, store and protect its contents from damage consistent with typical hospital or surgical center circulation (i.e. between central services and the operating room) and during transport between a Zimmer distributorship and a health care facility by Zimmer distributorship personnel.
The firm discovered that the current Personal Tibial/General Fixed Bracket Trays may not have the specific geometry to accommodate the tibial drill and stop guide with proper position/orientation. The drill bracket and drill guide are not the correct geometry to allow for a proper fit in the Persona Fixed Case Kit which could lead to improper sterilization of the tibial drill and stop guide.
Item Number 00-5907-015-00, Lot Numbers: 56518863 & 56518862
Worldwide Distribution - USA including AZ, CA, CO, FL, IL, IN, KS, MI, MN, NC, NJ, NY, OH, OR, PA, TN, TX & VA and Internationally to AUSTRALIA, FRANCE, INDIA, ITALY, and GERMANY.
190 units