FDA Enforcement Class II Terminated

QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.

Recall: Z-2284-2012 · Reported September 5, 2012

Enforcement

Recall Number
Z-2284-2012
Event ID
62776
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Steris Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Visit
Report Date
September 5, 2012
Initiation Date
August 10, 2012
Classification Date
August 28, 2012
Termination Date
July 5, 2013
Address
5960 Heisley Rd, N/A, Mentor, OH, 44060-1834, United States

Description

QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.

Reason

STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..

Code Info

Lot numbers: 1879717, 1943786, 2054914, 2143717, 2177723, 2342319, 2730752, 2922565, 3094190

Distribution

Nationwide distribution: USA including states of: AR, AK, CA, CO, DE, FL, IL, MA, MD, NC, NE, NH, NY, OH, PA, SD, TN, TX, WI, and WV.

Quantity

64 units