FDA Enforcement Class II Terminated

The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that is intended for medical purposes. It can be operated by the patient or the caregiver to adjust the height and surface contour of the bed. The bed includes movable and latchable siderails. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms.

Recall: Z-2283-2013 · Reported October 2, 2013

Enforcement

Recall Number
Z-2283-2013
Event ID
65922
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Medical Division of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 2, 2013
Initiation Date
August 19, 2013
Classification Date
September 26, 2013
Termination Date
January 29, 2014
Address
3800 E Centre Ave, N/A, Portage, MI, 49002-5826, United States

Description

The FL23SE Bed is an AC-powered adjustable hospital bed with four built-in electric DC motors and remote controls that is intended for medical purposes. It can be operated by the patient or the caregiver to adjust the height and surface contour of the bed. The bed includes movable and latchable siderails. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms. The FL23SE Bed is intended for use with patients for procedures, therapy, recovery in healthcare environments, and to transport patients between bays and procedural rooms.

Reason

The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.

Code Info

Model Numbers: FL23SE with nurse call option: OL235017 & OL235018 List of Affected Devices: J60267 J60425 J60009 J60001 J60094 J60095 J60096 J60348 J60349 J60350 J60351 J60352 J60353 J60354 J60360 J60361 J60362 J60363 J60364 J60365 J60366 J60367 J60368 J60369 J60370 J60371 J60372 J60373 J60374 J60375 J60376 J60377 J60378 J60379 J60380 J60381 J60382 J60383 J60384 J60385 J60386 J60387 J60388 J60389 J60390 J60391 J60392 J60393 J60394 J60395 J60396 J60397 J60398 J60399 J60400 J60401 J60402 J60403 J60404 J60405 J60512

Distribution

Worldwide Distribution - USA including MI, WI, VA and internationally to Canada.

Quantity

61 units