FDA Enforcement
Class II
Ongoing
Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System
Recall: Z-2282-2026
·
Reported June 10, 2026
Enforcement
- Recall Number
- Z-2282-2026
- Event ID
- 98894
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Stryker Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 10, 2026
- Initiation Date
- May 6, 2026
- Classification Date
- June 1, 2026
- Address
- 1941 Stryker Way, Portage, MI, 49002-9711, United States
Description
Neptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System
Reason
It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.
Code Info
GTIN 07613327634914, Serial Number 2521008873
Distribution
US Nationwide distribution in the state of Indiana.
Quantity
1 unit