Aries Bordetella Assay, REF 50-10037, UDI # 00840487101452
Enforcement
- Recall Number
- Z-2282-2019
- Event ID
- 83470
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Luminex Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 21, 2019
- Initiation Date
- February 20, 2019
- Classification Date
- August 15, 2019
- Address
- 12212 Technology Blvd, N/A, Austin, TX, 78727-6101, United States
Description
Aries Bordetella Assay, REF 50-10037, UDI # 00840487101452
Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
Lot numbers and discontinue use by date for Part Number 50-10037: AA7125 - Immediately, AA7310 - Immediately, AA7548 - 3/13/2019, AA7633 - 4/12/2019
US: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
N/A