FDA Enforcement Class II Terminated

Atellica CH Fructosamine Reagent (FRUC), Siemens Material Number 11097637, UDI Number 00630414595580 use in the quantitative measurement of glycated protein (fructosamine) in human serum and plasma for monitoring the blood glucose levels of diabetic patients over the preceding two-three weeks.

Recall: Z-2281-2020 · Reported June 17, 2020

Enforcement

Recall Number
Z-2281-2020
Event ID
85602
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 17, 2020
Initiation Date
May 4, 2020
Classification Date
June 5, 2020
Termination Date
September 20, 2021
Address
500 Gbc Dr, N/A, Newark, DE, 19702-2466, United States

Description

Atellica CH Fructosamine Reagent (FRUC), Siemens Material Number 11097637, UDI Number 00630414595580 use in the quantitative measurement of glycated protein (fructosamine) in human serum and plasma for monitoring the blood glucose levels of diabetic patients over the preceding two-three weeks.

Reason

The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

Code Info

Lot # 291011 ***Added 9/9/20*** 201115

Distribution

Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.

Quantity

26