FDA Enforcement Class II Terminated

Randox Liquid Assayed Specific Protein Control - Level 3 (SP Control 3), Cat. no. PS2684

Recall: Z-2280-2018 · Reported July 4, 2018

Enforcement

Recall Number
Z-2280-2018
Event ID
80211
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Randox Laboratories Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 4, 2018
Initiation Date
May 22, 2018
Classification Date
June 22, 2018
Termination Date
November 17, 2020
Address
55 Diamond Road, Crumlin (North), N/A, Ireland

Description

Randox Liquid Assayed Specific Protein Control - Level 3 (SP Control 3), Cat. no. PS2684

Reason

The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains.

Code Info

All lots.

Distribution

Distributed to accounts in CA, IN, KS, ME, and WV, and PR. Foreign distribution to Australia, Bangladesh, Belarus, Belgium, Brazil, Cambodia, Canada, Chile, China, Colombia, Cyprus, Denmark, Dutch Caribbean, Ecuador, Egypt, El Salvador, Ethiopia, Falkland Islands, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Iraq, Ireland, Italy, Jordan, Kenya, Saudi Arabia, Korea, Kuwait, Latvia, Lebanon, Peru, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Morocco, Myanmar, New Zealand, Norway, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Oman, Thailand, Netherlands, Turkey, United Arab Emirates, United Kingdom, Ukraine, Vietnam, and Zimbabwe.