FDA Enforcement Class II Ongoing

Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with OEC Flexiview 8800, OEC 9800, and OEC 9900 Systems

Recall: Z-2279-2023 · Reported August 9, 2023

Enforcement

Recall Number
Z-2279-2023
Event ID
92504
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
GE OEC Medical Systems, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 9, 2023
Initiation Date
May 15, 2023
Classification Date
July 31, 2023
Address
384 N Wright Brothers Dr, N/A, Salt Lake City, UT, 84116-2862, United States

Description

Field Upgrade Kits: P/N 5451208 RIU, 9 INCH, TOSHIBA; P/N 5485304 SERVICE PART, IMAGE INTENSIFIER, 9 INCH, TOSHIBA; P/N 5761613 SERVICE PART, IMAGE INTENSIFIER, 9 INCH - used with OEC Flexiview 8800, OEC 9800, and OEC 9900 Systems

Reason

Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.

Code Info

P/N/Sales Order Number: 5451208/80165102, 80171955, 80173514, 80162453, 80183950, 80151733, 80163547, 80171532, 80156410, 80176650, 10530514, 105449333, 80167595; 5485304/10557644, 118687327; 5761613/10283014, 10152211, 105108003, 105114645. OEC 9800, UDI-DI: 00840682114349; OEC 9900 Systems, UDI-DI: 00840682114301

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.

Quantity

16