OEC 9800 Systems with 9-inch Image Intensifier
Enforcement
- Recall Number
- Z-2276-2023
- Event ID
- 92504
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- GE OEC Medical Systems, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 9, 2023
- Initiation Date
- May 15, 2023
- Classification Date
- July 31, 2023
- Address
- 384 N Wright Brothers Dr, N/A, Salt Lake City, UT, 84116-2862, United States
Description
OEC 9800 Systems with 9-inch Image Intensifier
Image intensifiers on systems, used to provide fluoroscopic/digital spot images can become detached if the systems encounter a large impact force when moved, which may result in the image intensifier falling, which could lead to injuries such as fractures, hematoma, contusion, bruise, pain.
UDI-DI: 00840682114349. Serial Numbers: 89-0317, 89-0554, 8S-1293, 8S-0509, 8S-0108, 8S-1029, 8S-2836, 89-3028, 8S-1574, 89-0609-RC, 89-1308, 8S-0394, 89-0979, 8SXXXX03626, 8SXXXX03683, 8SXXXX03681, 8S-1894, 8S-0175, 8S-2988, 89-2459, 89-2101, 89-3245, 89-1623, 89-0296, 8S-0200, 8S-0748, 89-1590, 89-1713, 8S-0124, 89-0282, 8S-2580, 89-3132, 8S-3039, 89-3451, 89-0223, 89-1155, 8SXXXX03680, 8S-3278, 8S-2426, 89-2913, 8S-1167, 89-1278, 89-1854, 8S-0233-C, 8S-1926, 8SXXXX03682, 89-2167, 8S-1109, 8SXXXX03679, 8S-0410, 89-1579-N, 89-1920, 89-3102, 8S-2046-C, 89-2340, 89-0461, 89-3809, 89-3810
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Canada, Chile, Colombia, Czechia, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, France, Georgia, Germany, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, KOREA, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, New Caledonia, New Zealand, Nigeria, Paraguay, Peru, Poland, Portugal, Romania, Russia, Singapore, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
58