FDA Enforcement Class II Ongoing

PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

Recall: Z-2275-2026 · Reported June 10, 2026

Enforcement

Recall Number
Z-2275-2026
Event ID
98959
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Oculus Technologies of Mexico, S.A. de C.V.
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
June 10, 2026
Initiation Date
April 27, 2026
Classification Date
June 1, 2026
Address
Calle Industria Vidriera 81, Industrial Zapopan Norte, Zapopan, N/A, Mexico

Description

PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

Reason

The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.

Code Info

UDI:00860009268616 Lot: 25G263

Distribution

US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.

Quantity

5904 units