FDA Enforcement Class II Terminated

Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.

Recall: Z-2274-2016 · Reported August 3, 2016

Enforcement

Recall Number
Z-2274-2016
Event ID
74567
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 3, 2016
Initiation Date
February 5, 2016
Classification Date
July 22, 2016
Termination Date
January 17, 2017
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.

Reason

A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.

Code Info

AFB13H005, AFB13H008, AFB13H012, AFB13K002, AFB13K005, AFB13K008, AFB13J002, AFB13J003, AFB13J004, AFB13J005, AFB13J007, AFB15J001, AFB15C010, AFB13K009, AFB14H001, AFB15C008, AFB15C003, AFB15C005, AFB15C004, AFB15C006, AFB14H005, AFB14H003, AFB13K004, AFB13K006, AFB13K007, AFB13J008, AFB13K001, AFB13J006, AFB13J010, AFB13H013, AFB13H009, AFB13H003, AFB13H004

Distribution

Worldwide Distribution - US Nationwide (including Puerto Rico) and the countries of: Australia, Canada, Colombia, Germany, Guatemala, Netherlands, Singapore, Tunisia, , Spain

Quantity

62 units total (33 units in US)