Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.
Enforcement
- Recall Number
- Z-2274-2016
- Event ID
- 74567
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 3, 2016
- Initiation Date
- February 5, 2016
- Classification Date
- July 22, 2016
- Termination Date
- January 17, 2017
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
Allegra X-5 rotor yoke, Catalog No. (Instruments) B30589 and B30590 (Rotor) B30593. Product Labeling: IFU B29071AA. Product Usage: For In Vitro Diagnostic Use. The Allegra X-5 is intended for the separation of components through the use of relative centrifugal force. It is designed to separate human samples, including blood and other body fluids, for processing, analysis and in vitro diagnostic testing, as well as non-human body samples and chemicals, including industrial and environmental samples. This centrifuge should be operated by qualified personnel only.
A defective rotor yoke on Allegra X-5 allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.
AFB13H005, AFB13H008, AFB13H012, AFB13K002, AFB13K005, AFB13K008, AFB13J002, AFB13J003, AFB13J004, AFB13J005, AFB13J007, AFB15J001, AFB15C010, AFB13K009, AFB14H001, AFB15C008, AFB15C003, AFB15C005, AFB15C004, AFB15C006, AFB14H005, AFB14H003, AFB13K004, AFB13K006, AFB13K007, AFB13J008, AFB13K001, AFB13J006, AFB13J010, AFB13H013, AFB13H009, AFB13H003, AFB13H004
Worldwide Distribution - US Nationwide (including Puerto Rico) and the countries of: Australia, Canada, Colombia, Germany, Guatemala, Netherlands, Singapore, Tunisia, , Spain
62 units total (33 units in US)