FDA Enforcement Class II Ongoing

ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.

Recall: Z-2273-2023 · Reported August 9, 2023

Enforcement

Recall Number
Z-2273-2023
Event ID
92446
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Maquet Cardiovascular, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 9, 2023
Initiation Date
June 28, 2023
Classification Date
July 31, 2023
Address
45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States

Description

ACROBAT V Vacuum Off-Pump System, Part Number C-OM-9100S. Component of a chest stabilization system for cardiac and chest surgery.

Reason

The Standard Blade may not securely latch onto the Activator Drive. This may result in the inability to stabilize and position or may result in the loss of stabilization and positioning of heart, release of device component into the patient, and/or a procedural delay and/or conversion.

Code Info

UDI-DI: 00607567700536

Distribution

Worldwide distribution - US Nationwide and the countries of Bangladesh, Brazil, Chile, Colombia, Iran, Jordan, Malaysia, Mexico, Netherlands, Singapore, Sudan, Taiwan, Thailand, and United Arab Emirates.

Quantity

594 units