FDA Enforcement Class II Terminated

VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Unique Device Identifier No. 10758750009114 or 20758750009111, IVD -- Sales Unit: 300 slides/pack Product Usage: For in vitro diagnostic use only. VITROS Ca Slides quantitatively measure calcium (Ca) concentration in serum, plasma, and urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone disorders, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Recall: Z-2273-2016 · Reported August 3, 2016

Enforcement

Recall Number
Z-2273-2016
Event ID
74355
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho-Clinical Diagnostics
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 3, 2016
Initiation Date
May 18, 2016
Classification Date
July 22, 2016
Termination Date
June 8, 2018
Address
513 Technology Blvd, N/A, Rochester, NY, 14626-3601, United States

Description

VITROS Chemistry Products Ca Slides, GEN 29, REF/Product Code 145 0261, Unique Device Identifier No. 10758750009114 or 20758750009111, IVD -- Sales Unit: 300 slides/pack Product Usage: For in vitro diagnostic use only. VITROS Ca Slides quantitatively measure calcium (Ca) concentration in serum, plasma, and urine. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone disorders, chronic renal disease and tetany (intermittent muscular contractions or spasms).

Reason

Ortho Clinical Diagnostics (OCD) confirmed a potential for biased calcium (Ca) results to be generated when using VITROS Ca Slides, GEN29, and the VITROS 250/350/950/5,1 FS, 4600 Chemistry System or the VITROS 5600 Integrated System. An increase in negatively biased calcium results complaints prompted the investigation that identified this product issue.

Code Info

Lots #s: 0329-0528-8193 (exp. 05/01/2017), 0329-0528-8892 (exp. 06/01/2017), 0329-0528-8192 (exp. 05/01/2017), 0329-0528-8902 (exp. 06/01/2017), 0329-0528-8226 (exp. 05/01/2017), 0329-0528-8928 (exp. 07/01/2017), 0329-0528-8227 (exp. 05/01/2017), 0329-0528-9000 (exp. 07/01/2017), 0329-0528-8359 (exp. 05/01/2017), 0329-0528-8927 (exp. 07/01/2017), 0329-0528-8361 (exp. 05/01/2017), 0329-0528-8965 (exp. 07/01/2017), 0329-0528-8360 (exp. 05/01/2017), 0329-0528-8966 (exp. 07/01/2017), 0329-0528-8324 (exp. 05/01/2017), 0329-0528-9167 (exp. 07/01/2017), 0329-0528-8323 (exp. 05/01/2017), 0329-0528-9171 (exp. 07/01/2017), 0329-0528-8325 (exp. 05/01/2017), 0329-0528-9196 (exp. 07/01/2017), 0329-0528-6999 (exp. 03/01/2017), 0329-0528-9172 (exp. 07/01/2017), 0329-0528-7000 (exp. 03/01/2017), 0329-0528-9177 (exp. 07/01/2017), 0329-0528-6998 (exp. 03/01/2017), 0329-0528-9265 (exp. 07/01/2017), 0329-0528-8903 (exp. 06/01/2017), 0329-0528-9176 (exp. 07/01/2017), 0329-0528-8901 (exp. 06/01/2017), 0329-0528-9264 (exp. 07/01/2017), 0329-0528-8893 (exp. 06/01/2017), 0329-0528-9368 (exp. 07/01/2017), 0329-0528-8891 (exp. 06/01/2017), 0329-0528-9369 (exp. 07/01/2017).

Distribution

Worldwide Distribution - US Nationwide in the countries of Puerto Rico and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, Poland and Venezuela.

Quantity

US: 27,625 sales units; Foreign: 4757 sales units