FDA Enforcement Class II Terminated

TRESTLE LUXE¿ Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.

Recall: Z-2273-2012 · Reported September 5, 2012

Enforcement

Recall Number
Z-2273-2012
Event ID
51437
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Alphatec Spine, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 5, 2012
Initiation Date
May 29, 2012
Classification Date
August 27, 2012
Termination Date
September 4, 2012
Address
5818 El Camino Real, N/A, Carlsbad, CA, 92008-8816, United States

Description

TRESTLE LUXE¿ Anterior Cervical Plating System, Product code KWQ. 510(k) K102820. The TRESTLE LUXE Anterior Cervical Plating System is a temporary device used to stabilize the cervical spine during bone fusion development. Device implants include a range of plate sizes and bone screws to provide the versatility required for the specific indications noted. Fixation is achieved by means of a rigid plate that is surgically attached to the spine with bone screws. The TRESTLE LUXE Anterior Cervical Plating System is intended for use in the anterior cervical spine (C2-C7). Patients with trauma (including fractures), spondylolisthesis, and pseudoarthrosis, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spinal stenosis, tumors, and failed previous fusion.

Reason

On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.

Code Info

Product code KWQ. 510(k) K102820 Part Number 71721. Lot numbers: 6629701, 6594701, 6434301, 6434302, 6434302R1.

Distribution

Nationwide Distribution.

Quantity

38