FDA Enforcement Class II Ongoing

DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461

Recall: Z-2270-2025 · Reported August 13, 2025

Enforcement

Recall Number
Z-2270-2025
Event ID
97162
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 13, 2025
Initiation Date
June 25, 2025
Classification Date
August 5, 2025
Address
1000 Lake Hazeltine Dr, N/A, Chaska, MN, 55318-1037, United States

Description

DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461

Reason

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

Code Info

UDI-DI: 15099590369248; Serial Numbers: 902101

Distribution

Worldwide distribution - US Nationwide.

Quantity

1 system