FDA Enforcement
Class II
Ongoing
Always-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-5300
Recall: Z-2267-2023
·
Reported August 9, 2023
Enforcement
- Recall Number
- Z-2267-2023
- Event ID
- 92383
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Olympus Corporation of the Americas
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 9, 2023
- Initiation Date
- May 15, 2023
- Classification Date
- July 28, 2023
- Address
- 3500 Corporate Pkwy, N/A, Center Valley, PA, 18034-8229, United States
Description
Always-On Tip Tracked Instruments (SPiN Drive instruments) Triple Needle Brush, 12mm L, 1.8mm OD-Intended for use in clinical interventions and for anatomical structures where computed tomography and/or endoscopic brINS-0362onchoscopy are currently used for visualizing Model: INS-5300
Reason
Electromagnetic (EM) sensor tracking malfunction impacts the use of navigation functionality during endobronchial procedures and prompts the system to report a coil break warning message. This could result in a potential procedural delay
Code Info
UDI-DI:00815686021207 Lot Number: 04617200805 and below
Distribution
Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: CN, DE, HK, IT, KR, SG, TH & TW.
Quantity
759.8 units