FDA Enforcement Class II Terminated

IntelePACS software versions 3-8-1 to 4-9-1, Product Usage: IntelePACS is a device that receives digital images and data from various sources (such as, CT scanners, MRI scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources.

Recall: Z-2267-2014 · Reported August 27, 2014

Enforcement

Recall Number
Z-2267-2014
Event ID
68719
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Intelerad Medical Systems
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 27, 2014
Initiation Date
July 3, 2014
Classification Date
August 19, 2014
Termination Date
January 13, 2016
Address
895 Rue De La Gauchetiere O, Montreal, N/A, Canada

Description

IntelePACS software versions 3-8-1 to 4-9-1, Product Usage: IntelePACS is a device that receives digital images and data from various sources (such as, CT scanners, MRI scanners, ultrasound systems, R/F units, computer and direct radiographic devices, secondary capture devices, scanners, imaging gateways, or other imaging sources.

Reason

In specific situation a partially truncated file may be written to the IntelePACS server. This partially truncated file still can be viewed by InteleViewer and InteleViewer will report to the user that the file in question is complete. This situation only occurs if a file is partially truncated.

Code Info

software versions 3-8-1 to 4-9-1

Distribution

Worldwide Distribution - USA Nationwide and the countries of Scotland, Ireland, United Kingdom, Canada, Australia, and New Zealand.

Quantity

230