FDA Enforcement
Class II
Terminated
Medline Silicone Drain 15Fr 3/16" Full Fluted PN: DYNJWE2188HB An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
Recall: Z-2266-2019
·
Reported August 21, 2019
Enforcement
- Recall Number
- Z-2266-2019
- Event ID
- 83287
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Degania Silicone, Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 21, 2019
- Initiation Date
- June 28, 2019
- Classification Date
- August 12, 2019
- Termination Date
- October 22, 2020
- Address
- Degania Bet, N/A, Dganya Bet, N/A, N/A, Israel
Description
Medline Silicone Drain 15Fr 3/16" Full Fluted PN: DYNJWE2188HB An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
Reason
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
Code Info
Lot Code: P1841839
Distribution
Nationwide Distribution
Quantity
200