FDA Enforcement
Class III
Ongoing
Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA
Recall: Z-2265-2025
·
Reported August 13, 2025
Enforcement
- Recall Number
- Z-2265-2025
- Event ID
- 97196
- Classification
- Class III
- Status
- Ongoing
- Product Type
- Devices
- Firm
- DRG International, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 13, 2025
- Initiation Date
- June 4, 2025
- Classification Date
- August 4, 2025
- Address
- 841 Mountain Ave, N/A, Springfield, NJ, 07081-3437, United States
Description
Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA
Reason
An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration dates on the bottle labels were correct (i.e., 2026-01-31).
Code Info
Model/Catalog Number: EIA-4568; UDI-DI: 04048474045680; Lot number: 304K034;
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, PA and the countries of Germany, Hong Kong, Moldova.
Quantity
47 units