FDA Enforcement
Class II
Terminated
Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
Recall: Z-2265-2019
·
Reported August 21, 2019
Enforcement
- Recall Number
- Z-2265-2019
- Event ID
- 83287
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Degania Silicone, Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- August 21, 2019
- Initiation Date
- June 28, 2019
- Classification Date
- August 12, 2019
- Termination Date
- October 22, 2020
- Address
- Degania Bet, N/A, Dganya Bet, N/A, N/A, Israel
Description
Cardinal Health Jackson-Pratt Channel Drain 19FR, 3/4 flutes, with trocar Cardinal Health: PN JP-2225 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
Reason
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
Code Info
Lot Codes: P1851056 P1867059
Distribution
Nationwide Distribution
Quantity
2160