FDA Enforcement Class II Terminated

Synthes 3.0 MM TI Cannulated Screw Instrument and Implant Set

Recall: Z-2265-2014 · Reported August 27, 2014

Enforcement

Recall Number
Z-2265-2014
Event ID
68738
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Synthes, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 27, 2014
Initiation Date
June 30, 2014
Classification Date
August 18, 2014
Termination Date
December 18, 2015
Address
1302 Wrights Ln E, West Chester, PA, 19380-3417, United States

Description

Synthes 3.0 MM TI Cannulated Screw Instrument and Implant Set

Reason

A screw set was returned from the field for destruction after being exposed to flood water, which was erroneously reprocessed and returned to the field. Initial investigative testing indicated the components within the set may be cytotoxic.

Code Info

Part Number: 145.17E Lot Number: EV01609, individual part numbers contained within 145.17E: 304.49 309.501 310.221 310.804 311.43 312.151 312.153 313.969 314.463 314.464 314.465 319.291 319.292 319.702 319.97 398.408 398.409 402.608-402.640 402.714-402.740 419.89 419.972 292.622, and 292.623.

Distribution

Distributed in the states of WI and MI.

Quantity

4