Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended use environment includes anatomic pathology (AP) laboratories and histology facilities within hospital, clinic, reference, or private lab settings. The System is to be operated by qualified users in
Enforcement
- Recall Number
- Z-2262-2020
- Event ID
- 85526
- Classification
- Class II
- Status
- Completed
- Product Type
- Devices
- Firm
- Ventana Medical Systems Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 17, 2020
- Initiation Date
- October 4, 2019
- Classification Date
- June 5, 2020
- Address
- 1910 E Innovation Park Dr, N/A, Oro Valley, AZ, 85755-1962, United States
Description
Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended use environment includes anatomic pathology (AP) laboratories and histology facilities within hospital, clinic, reference, or private lab settings. The System is to be operated by qualified users in
The firm became aware of a cleaning solution leaking issue from the middle staining module of the Ventana HE 600 instrument, used for histologic section samples in anatomic pathology laboratory settings, which caused fluid to travel down to the electrical wiring connector of the Coverslipper solvent shield circuit board. This can cause an electrical short and result in burned components.
All lots
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MD, MI, MA, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Costa Rica, Croatia, Denmark, Ecuador, El Salvador, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Kuwait, Lebanon, Malaysia, Norway, Peru, Poland, Saudi Arabia, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.