FDA Enforcement
Class II
Ongoing
Medline medical convenience kit labeled as EYE TRAY-LF, Medline Kit SKU DYNJ21627P
Recall: Z-2260-2026
·
Reported June 3, 2026
Enforcement
- Recall Number
- Z-2260-2026
- Event ID
- 98866
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- June 3, 2026
- Initiation Date
- April 10, 2026
- Classification Date
- May 28, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline medical convenience kit labeled as EYE TRAY-LF, Medline Kit SKU DYNJ21627P
Reason
Specific Medline Kits contain Cardinal Health Monoject 1mL Luer Lock Syringes, Tuberculin Print, Sterile. Cardinal Health has recalled these syringes because the outer carton and blister pack are labeled as a 1 mL Luer Lock Tuberculin Syringe (Product Code 1180100777), but the syringes inside are U-100 insulin syringes.
Code Info
UDI/DI 10198459151569 (each), 40198459151560 (case), Lot Numbers: 26BMC538
Distribution
US Nationwide distribution in the states of California.
Quantity
8 kits