FDA Enforcement Class II Terminated

Cardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.

Recall: Z-2260-2019 · Reported August 21, 2019

Enforcement

Recall Number
Z-2260-2019
Event ID
83287
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Degania Silicone, Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 21, 2019
Initiation Date
June 28, 2019
Classification Date
August 12, 2019
Termination Date
October 22, 2020
Address
Degania Bet, N/A, Dganya Bet, N/A, N/A, Israel

Description

Cardinal Health Jackson-Pratt Channel Drain 15FR Cardinal PN:JP-2223 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.

Reason

15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body

Code Info

Lot Codes: 1900956, P1900955

Distribution

Nationwide Distribution

Quantity

640 units