MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-1511; NGP 640G 1.8ml (mg/dL), Model No. 1512; NGP 640G PLGM 3ml (mmol/L), Model No. MMT-1711; NGP 640G PLGM 3ml (mg/dL), Model No. MMT-1712.
Recall: Z-2260-2015
·
Reported August 5, 2015
Enforcement
- Recall Number
- Z-2260-2015
- Event ID
- 70897
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medtronic MiniMed Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 5, 2015
- Initiation Date
- June 19, 2015
- Classification Date
- July 24, 2015
- Termination Date
- October 2, 2015
- Address
- 18000 Devonshire St, N/A, Northridge, CA, 91325-1219, United States
Description
MiniMed NGP 640G 1.8ml (mmol/L), Model No. MMT-1511; NGP 640G 1.8ml (mg/dL), Model No. 1512; NGP 640G PLGM 3ml (mmol/L), Model No. MMT-1711; NGP 640G PLGM 3ml (mg/dL), Model No. MMT-1712.
Reason
Medtronic MiniMed is recalling the MiniMed 620G and 640G insulin pumps because there are certain scenarios where the set Bolus screen will not timeout, which could cause confusion by showing a bolus amount that is no longer appropriate.
Code Info
n/a
Distribution
Australia, Czech Republic, Denmark, Finland, France, Ireland, Netherlands, Norway, Sweden, United Kingdom.
Quantity
1936 units