Computed Tomography X-ray Systems Ingenuity CT scanners. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
Enforcement
- Recall Number
- Z-2259-2015
- Event ID
- 71124
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems (Cleveland) Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 5, 2015
- Initiation Date
- February 5, 2015
- Classification Date
- July 24, 2015
- Termination Date
- June 26, 2018
- Address
- 595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States
Description
Computed Tomography X-ray Systems Ingenuity CT scanners. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.
5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict
Model #728326; Serial #s: 30003, 30004, 30005, 30006, 30007, 30008, 30009, 30011, 30012, 30013, 30015, 30018, 30019, 30020, 30025, 32086, 32093, 32107, 52023, 52044, 52051, 300003, 300004, 300005, 300010, 300011, 300012, 300013, 300014, 300016, 300017, 300019, 300020, 300021, 300024, 300025, 300027, 300028, 300029, 300031, 300032, 300033, 300034, 300035, 300036, 300037, 300039, 300040, 300041, 300042, 300043, 300044, 300045, 300047, 300049, 300050, 300051, 300052, 300053, 300054, 300055, 300056, 300057, 300058, 300059, 300060, 300061, 300062, 300063, 300065, 300066, 300068, 300069, 300070, 300071, 300072, 300073, 300074, 300075, 300077, 300078, 300079, 300080, 300081, 300082, 300084, 300086, 300087, 300088, 300089, 300090, 300091, 300092, 300093, 300094, 300095, 300096, 300097, 300098, 300099, 300100, 300101, 300102, 300103, 300104, 300105, 300106, 300107, 300108, 300110, 300111, 300112, 300113, 300114, 300115, 300116, 300117, 300118, 300119, 300120, 300121, 300122, 300123, 300124, 300125, 300126, 300127, 300128, 300129, 300130, 300131, 300132, 300135, 300136, 300139, 300140, 300142, 300143, 300144, 300145, 300146, 300147, 300148, 300149, 300150, 300151, 300156, 300157, 300158, 300159, 300160, 300161, 300162, 300163, 300164, 300165, 300166, 310058 & 30004.
Worldwide Distribution-US (nationwide) including the states of AR, AZ, CA, CT, DE, FL, GA, IA, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VT, WV & WY, and to the following countries: Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Cambodia, Canada, China, Colombia, Denmark, Finland, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Italy, Japan, Latvia, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Panama, Philippines, Russia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Viet Nam.
159 units