FDA Enforcement Class II Terminated

TRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF 71675202, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 13MM X 18 CM NAIL, 125o, REF 71675203, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 20 CM NAIL, 125o, REF 71675204, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 20 CM NAIL, 125o, REF 71675205, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 13MM X 20 CM NAIL, 125o, REF 71675206, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 130o, REF 71675207, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 130o, REF 71675208, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 13MM X 18 CM NAIL, 130o, REF 71675209, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 20 CM NAIL, 130o, REF 71675210, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 20 CM NAIL, 130o, REF 71675211, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 13MM X 20 CM NAIL, 130o, REF 71675212, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 26 CM NAIL, 125o, LEFT, REF 71675213, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 32 CM NAIL, 125o, LEFT, REF 71675219, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 34 CM NAIL, 125o, LEFT, REF 71675221, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 34 CM NAIL, 125o, RIGHT, REF 71675222, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 36 CM NAIL, LEFT, 125o, REF 71675223, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 38 CM NAIL, 125o, LEFT, REF 71675225, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 40 CM NAIL, 125o, LEFT, REF 71675227, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 40 CM NAIL, 125o, RIGHT, REF 71675228, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 28 CM NAIL, 125o, RIGHT, REF 71675238, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 32 CM NAIL, 125o, LEFT, REF 71675241, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 34 CM NAIL, 125o, LEFT, REF 71675243, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 34 CM NAIL, 125o, RIGHT, REF 71675244, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 36 CM NAIL, 125o, LEFT, REF 71675245, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 36 CM NAIL, 125o, RIGHT, REF 71675246, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 38 CM NAIL, 125o, LEFT, REF 71675247, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 38 CM NAIL, 125o, RIGHT, REF 71675248, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 40 CM NAIL, 125o, LEFT, REF 71675249, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 40 CM NAIL, 125o, RIGHT, REF 71675250, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 26 CM NAIL, 130o, LEFT, REF 71675257, QTY (1), STERILE R, Smi

Recall: Z-2258-2014 · Reported August 27, 2014

Enforcement

Recall Number
Z-2258-2014
Event ID
68792
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Smith & Nephew, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 27, 2014
Initiation Date
July 2, 2014
Classification Date
August 17, 2014
Termination Date
August 20, 2019
Address
1450 E Brooks Rd, Memphis, TN, 38116-1804, United States

Description

TRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 125o, REF 71675201, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 125o, REF 71675202, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 13MM X 18 CM NAIL, 125o, REF 71675203, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 20 CM NAIL, 125o, REF 71675204, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 20 CM NAIL, 125o, REF 71675205, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 13MM X 20 CM NAIL, 125o, REF 71675206, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 18 CM NAIL, 130o, REF 71675207, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 18 CM NAIL, 130o, REF 71675208, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 13MM X 18 CM NAIL, 130o, REF 71675209, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 20 CM NAIL, 130o, REF 71675210, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 20 CM NAIL, 130o, REF 71675211, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 13MM X 20 CM NAIL, 130o, REF 71675212, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 26 CM NAIL, 125o, LEFT, REF 71675213, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 32 CM NAIL, 125o, LEFT, REF 71675219, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 34 CM NAIL, 125o, LEFT, REF 71675221, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 34 CM NAIL, 125o, RIGHT, REF 71675222, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 36 CM NAIL, LEFT, 125o, REF 71675223, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 38 CM NAIL, 125o, LEFT, REF 71675225, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 40 CM NAIL, 125o, LEFT, REF 71675227, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 40 CM NAIL, 125o, RIGHT, REF 71675228, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 28 CM NAIL, 125o, RIGHT, REF 71675238, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 32 CM NAIL, 125o, LEFT, REF 71675241, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 34 CM NAIL, 125o, LEFT, REF 71675243, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 34 CM NAIL, 125o, RIGHT, REF 71675244, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 36 CM NAIL, 125o, LEFT, REF 71675245, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 36 CM NAIL, 125o, RIGHT, REF 71675246, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 38 CM NAIL, 125o, LEFT, REF 71675247, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 38 CM NAIL, 125o, RIGHT, REF 71675248, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 40 CM NAIL, 125o, LEFT, REF 71675249, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 11.5MM X 40 CM NAIL, 125o, RIGHT, REF 71675250, QTY (1), STERILE R, Smith & Nephew, Memphis, Inc., TN 38116 TRIGEN(R) INTERTAN(TM) 10MM X 26 CM NAIL, 130o, LEFT, REF 71675257, QTY (1), STERILE R, Smi

Reason

There was an error in manufacturing which resulted in the proximal lag screw hole being positioned incorrectly in several batches of TRIGEN(R); INTERTAN(TM); NAILS.

Code Info

TRIGEN INTERTAN 10MM X 18 CM NAIL, 125o, REF 71675201, Lot Numbers: 14CM05726, 14DM08712, 14CM12004, 14DM08713, 14CM12005, 14DM08714, 14CM12006, 14DM08715, 14CM12008, 14DM09828, 14CM12009, 14DM09829, 14CM15434, 14DM09830, 14CM15435, 14DM11951, 14CM15438, 14DM11952, 14CM15488, 14DM16264, 14CM15490, 14DM19617, 71675201 14CM17964, 14CM17965, 14CM17966, 14CM17966, 14DM01455, 14DM01456, 14DM04231, 14DM04232, 14DM04233, 14DM04234, 14DM04235, 14DM05411, 14DM05412, 71675201 14DM05413, 14DM05414, 14DM08711 TRIGEN INTERTAN 11.5MM X 18 CM NAIL, 125o, REF 71675202, Lot Numbers: 14CM15413, 14CM15415, 14CM15416, 14CM15417, 14CM15418, 14DM06190, 14DM06191, 14DM06192, 14DM06193, 14DM09835, 14DM09837 TRIGEN INTERTAN 13MM X 18 CM NAIL, 125o, REF 71675203, Lot Numbers: 14CM15443, 14CM01161, 14CM03675, 14CM05731, 14CM05732 TRIGEN INTERTAN 10MM X 20 CM NAIL, 125o, REF 71675204, Lot Numbers: 14CM15480 TRIGEN INTERTAN 11.5MM X 20 CM NAIL, 125o, REF 71675205, Lot Numbers: 14CM16786 TRIGEN INTERTAN 13MM X 20 CM NAIL, 125o, REF 71675206, Lot Numbers: 14CM03676, 14CM15471, 14CM16787 TRIGEN INTERTAN 10MM X 18 CM NAIL, 130o, REF 71675207, Lot Numbers: 14DM06197, 14DM08716, 14DM13210, 14DM13212 TRIGEN INTERTAN 11.5MM X 18 CM NAIL, 130o, REF 71675208, Lot Numbers: 14DM01460, 14DM01461, 14DM01462, 14DM01463, 14DM01464 TRIGEN INTERTAN 13MM X 18 CM NAIL, 130o, REF 71675209, Lot Numbers: 14CM15467, 14CM16788, 14CM16789, 14DM11044 TRIGEN INTERTAN 10MM X 20 CM NAIL, 130o, REF 71675210, Lot Numbers: 14CM15466, 14CM18995, 14CM18996, 14CM18997, 14CM18998, 14DM00764, 14DM00765, 14DM00766, 14DM00767 TRIGEN INTERTAN 11.5MM X 20 CM NAIL, 130o, REF 71675211, Lot Numbers: 14CM15464, 14CM18999, 14CM19000, 14CM21712, 14CM19001, 14CM21720, 14DM00769, 14DM00770 TRIGEN INTERTAN 13MM X 20 CM NAIL, 130o, REF 71675212, Lot Numbers: 14CM15463, 14CM16790 TRIGEN INTERTAN 10MM X 26 CM NAIL, 125o, LEFT, REF 71675213, Lot Numbers: 14DM14689 TRIGEN INTERTAN 10MM X 32 CM NAIL, 125o, LEFT, REF 71675219, Lot Numbers: 14DM17741 TRIGEN INTERTAN 10MM X 34 CM NAIL, 125o, LEFT, REF 71675221, Lot Numbers: 14DM08628 TRIGEN INTERTAN 10MM X 34 CM NAIL, 125o, RIGHT, REF 71675222, Lot Numbers: 14CM20693, 14DM16827 TRIGEN INTERTAN 10MM X 36 CM NAIL, LEFT, 125o, REF 71675223, Lot Numbers: 14CM19633 TRIGEN INTERTAN 10MM X 38 CM NAIL, 125o, LEFT, REF 71675225, Lot Numbers: 14CM17883, 14CM17884, 14CM17885, 14CM17886, 14DM04200, 14DM04201 TRIGEN INTERTAN 10MM X 40 CM NAIL, 125o, LEFT, REF 71675227, Lot Numbers: 14DM02910 TRIGEN INTERTAN 10MM X 40 CM NAIL, 125o, RIGHT, REF 71675228, Lot Numbers: 14DM11022 TRIGEN INTERTAN 11.5MM X 28 CM NAIL, 125o, RIGHT, REF 71675238, Lot Numbers: 14DM07429 TRIGEN INTERTAN 11.5MM X 32 CM NAIL, 125o, LEFT, REF 71675241, Lot Numbers: 14DM04202, 14CM10084 TRIGEN INTERTAN 11.5MM X 34 CM NAIL, 125o, LEFT, REF 71675243, Lot Numbers: 14CM21692 TRIGEN INTERTAN 11.5MM X 34 CM NAIL, 125o, RIGHT, REF 71675244, Lot Numbers: 14DM05253, 14DM05254 TRIGEN INTERTAN 11.5MM X 36 CM NAIL, 125o, LEFT, REF 71675245, Lot Numbers: 14DM14008 TRIGEN INTERTAN 11.5MM X 36 CM NAIL, 125o, RIGHT, REF 71675246, Lot Numbers: 14DM02912, 14DM14691 TRIGEN INTERTAN 11.5MM X 38 CM NAIL, 125o, LEFT, REF 71675247, Lot Numbers: 14DM05255 TRIGEN INTERTAN 11.5MM X 38 CM NAIL, 125o, RIGHT, REF 71675248, Lot Numbers: 14DM02913, 14DM02913, 14DM04204, 14DM04205 TRIGEN INTERTAN 11.5MM X 40 CM NAIL, 125o, LEFT, REF 71675249, Lot Numbers: 14DM07434, 14CM16262 TRIGEN INTERTAN 11.5MM X 40 CM NAIL, 125o, RIGHT, REF 71675250, Lot Numbers: 14DM11931 TRIGEN INTERTAN 10MM X 26 CM NAIL, 130o, LEFT, REF 71675257, Lot Numbers: 14CM14081 TRIGEN INTERTAN 10MM X 26 CM NAIL, 130o, RIGHT, REF 71675258, Lot Numbers: 14CM17967 TRIGEN INTERTAN 10MM X 30 CM NAIL, 130o, LEFT, REF 71675261, Lot Numbers: 14CM10088 TRIGEN INTERTAN 10MM X 30 CM NAIL, 130o, RIGHT, REF 71675262, Lot Numbers: 14CM15830 TRIGEN INTERTAN 10MM X 32 CM NAIL, 130o, LEFT, REF 71675263, Lot Numbers: 14CM18422, 14CM21682 TRIGEN INTERTAN 10MM X 32 CM NAIL, 130o, RIGHT, REF 71675264, Lot Numbers: 14CM15833, 14CM15834 TRIGEN INTERTAN 10MM X 34 CM NAIL, 130o, LEFT, REF 71675265, Lot Numbers: 14CM20698, 14CM20700, 14DM16204 TRIGEN INTERTAN 10MM X 34 CM NAIL, 130o, RIGHT, REF 71675266, Lot Numbers: 14CM15837, 14DM11023, 14DM11024, 14DM11025 TRIGEN INTERTAN 10MM X 36 CM NAIL, 130o, LEFT, REF 71675267, Lot Numbers: 14CM21686, 14DM03076, 14DM16823 TRIGEN INTERTAN 10MM X 36 CM NAIL, 130o, RIGHT, REF 71675268, Lot Numbers: 14CM15839, 14DM11932, 14DM11933 TRIGEN INTERTAN 10MM X 38 CM NAIL, 130o, LEFT, REF 71675269, Lot Numbers: 14CM10090 TRIGEN INTERTAN 10MM X 38 CM NAIL, 130o, RIGHT, REF 71675270, Lot Numbers: 14DM07441 TRIGEN INTERTAN 10MM X 40 CM NAIL, 130o, RIGHT, REF 71675272, Lot Numbers: 14CM10091 TRIGEN INTERTAN 10MM X 42 CM NAIL, 130o, LEFT, REF 71675273, Lot Numbers: 14CM10092 TRIGEN INTERTAN 11.5MM X 26 CM NAIL, 130o, LEFT, REF 71675279 Lot Numbers: 14DM08119 TRIGEN INTERTAN 11.5MM X 26 CM NAIL, 130o, RIGHT, REF 71675280 Lot Numbers: 14DM07383 TRIGEN INTERTAN 11.5MM X 28 CM NAIL, 130o, RIGHT, REF 71675282, Lot Numbers: 14DM07474 TRIGEN INTERTAN 11.5MM X 30 CM NAIL, 130o, RIGHT, REF 71675284, Lot Numbers: 14DM07480 TRIGEN INTERTAN 11.5MM X 32 CM NAIL, 130o, RIGHT, REF 71675286, Lot Numbers: 14DM07465, 14DM07468 TRIGEN INTERTAN 11.5MM X 34 CM NAIL, 130o, LEFT, REF 71675287, Lot Numbers: 14CM10093, 14DM07379 TRIGEN INTERTAN 11.5MM X 34 CM NAIL, 130o, RIGHT, REF 71675288, Lot Numbers: 14CM10094, 14DM07373 TRIGEN INTERTAN 11.5MM X 36 CM NAIL, 130o, RIGHT, REF 71675290, Lot Numbers: 14CM10095, 14DM07457 TRIGEN INTERTAN 11.5MM X 38 CM NAIL, 130o, LEFT, REF 71675291, Lot Numbers: 14CM09471 TRIGEN INTERTAN 11.5MM X 38 CM NAIL, 130o, RIGHT, REF 71675292, Lot Numbers: 14CM10097, 14DM13214 TRIGEN INTERTAN 11.5MM X 40 CM NAIL, 130o, RIGHTT, REF 71675294, Lot Numbers: 14DM07448 TRIGEN INTERTAN 10S 10MM X 18 CM NAIL, 125o, REF 71675383, Lot Numbers: 14CM08470, 14CM08471, 14CM08472, 14CM08473, 14CM08474, 14CM15447, 14CM15449, 14CM15451, 14CM20793, 14CM20799, 14DM02914, 14DM02917, 14DM02918, 14DM13216, 14DM13217, 14DM14696, 14DM14697, 14DM14698 TRIGEN INTERTAN 10S 10MM X 18 CM NAIL, 130o, REF 71675384, Lot Numbers: 14CM20800, 14CM20803, 14DM02919, 14DM02920, 14DM02921, 14DM04240, 14DM04241, 14DM05416 TRIGEN INTERTAN 10S 10MM X 20 CM NAIL, 125o, REF 71675385, Lot Numbers: 14DM19439 TRIGEN INTERTAN 13MM X 38 CM NAIL, 125o, LEFT, REF 71675413, Lot Numbers: 14CM16791 TRIGEN INTERTAN 1.5 10MM X 26 CM NAIL, 125o, LEFT, REF 71676501, Lot Numbers: 14CM17968 TRIGEN INTERTAN 1.5 10MM X 30 CM NAIL, 125o, RIGHT, REF 71676506, Lot Numbers: 14DM14702 TRIGEN INTERTAN 1.5 10MM X 34 CM NAIL, 125o, RIGHT, REF 71676510, Lot Numbers: 14CM10739 TRIGEN INTERTAN 1.5 10MM X 40 CM NAIL, 125o, LEFT, REF 71676515, Lot Numbers: 14CM10740 TRIGEN INTERTAN 1.5 10MM X 44 CM NAIL, 125o, RIGHT, REF 71676520, Lot Numbers: 14CM20366, 14DM16205 TRIGEN INTERTAN 1.5 10MM X 36 CM NAIL, 130o, RIGHT, REF 71676534, Lot Numbers: 14DM08632 TRIGEN INTERTAN 1.5 10MM X 38 CM NAIL, 130o, LEFT, REF 71676535, Lot Numbers: 14CM16762, 14CM16763 TRIGEN INTERTAN 1.5 10MM X 44 CM NAIL, 130o, LEFT, REF 71676541, Lot Numbers: 14CM19634 TRIGEN INTERTAN 1.5 11.5MM X 28 CM NAIL, 125o, LEFT, REF 71676547, Lot Numbers: 14DM01466 TRIGEN INTERTAN 1.5 11.5MM X 28 CM NAIL, 125o, RIGHT, REF 71676548, Lot Numbers: 14DM05257 TRIGEN INTERTAN 1.5 11.5MM X 30 CM NAIL, 125o, LEFT, REF 71676549, Lot Numbers: 14DM01467 TRIGEN INTERTAN 1.5 11.5MM X 30 CM NAIL, 125o, RIGHT, REF 71676550, Lot Numbers: 14DM01468 TRIGEN INTERTAN 1.5 11.5MM X 32 CM NAIL, 125o, RIGHT, REF 71676552, Lot Numbers: 14DM13198 TRIGEN INTERTAN 1.5 11.5MM X 34 CM NAIL, 125o, LEFT, REF 71676553, Lot Numbers: 14CM18424, 14DM14296, 14DM17751, 14DM17754 TRIGEN INTERTAN 1.5 11.5MM X 34 CM NAIL, 125o, RIGHT, REF 71676554, Lot Numbers: 14CM21694, 14DM01470 TRIGEN INTERTAN 1.5 11.5MM X 36 CM NAIL, 125o, LEFT, REF 71676555, Lot Numbers: 14CM21690, 14DM17740 TRIGEN INTERTAN 1.5 11.5MM X 36 CM NAIL, 125o, RIGHT, REF 71676556, Lot Numbers: 14CM05007, 14DM11026 TRIGEN INTERTAN 1.5 11.5MM X 38 CM NAIL, 125o, LEFT, REF 71676557, Lot Numbers: 14DM09815 TRIGEN INTERTAN 1.5 11.5MM X 38 CM NAIL, 125o, RIGHT, REF 71676558, Lot Numbers: 14DM02922, 14DM09816, 14DM09817 TRIGEN INTERTAN 1.5 11.5MM X 40 CM NAIL, 125o, LEFT, REF 71676559, Lot Numbers: 14DM09819 TRIGEN INTERTAN 1.5 11.5MM X 40 CM NAIL, 125o, RIGHT, REF 71676560, Lot Numbers: 14DM00771, 14DM00772, 14DM00773, 14DM00774, 14DM00775, 14DM00776, 14DM17748 TRIGEN INTERTAN 1.5 11.5MM X 42 CM NAIL, 125o, LEFT, REF 71676561, Lot Numbers: 14CM19696, 14CM19698, 14CM19699, 14CM19703 TRIGEN INTERTAN 1.5 11.5MM X 42 CM NAIL, 125o, RIGHT, REF 71676561, Lot Numbers: 14CM19703 TRIGEN INTERTAN 1.5 11.5MM X 44 CM NAIL, 125o, LEFT, REF 71676563, Lot Numbers: 14DM16824 TRIGEN INTERTAN 1.5 11.5MM X 44 CM NAIL, 125o, RIGHT, REF 71676564, Lot Numbers: 14CM20701, 14CM20703 TRIGEN INTERTAN 1.5 11.5MM X 46 CM NAIL, 125o, LEFT, REF 71676565, Lot Numbers: 14DM05259 TRIGEN INTERTAN 1.5 11.5MM X 26 CM NAIL, 130o, RIGHT, REF 71676568, Lot Numbers: 14CM20308 TRIGEN INTERTAN 1.5 11.5MM X 28 CM NAIL, 130o, RIGHT, REF 71676570, Lot Numbers: 14CM11204 TRIGEN INTERTAN 1.5 11.5MM X 30 CM NAIL, 130o, RIGHT, REF 71676572, Lot Numbers: 14DM06198 TRIGEN INTERTAN 1.5 11.5MM X 32 CM NAIL, 130o, LEFT, REF 71676573, Lot Numbers: 11JM10220B, 14CM20310, 14CM20311, 14DM06199 TRIGEN INTERTAN 1.5 11.5MM X 32 CM NAIL, 130o, RIGHT, REF 71676574, Lot Numbers: 14CM14066, 14DM06200, 14DM06202 TRIGEN INTERTAN 1.5 11.5MM X 34 CM NAIL, 130o, LEFT, REF 71676575, Lot Numbers: 11FM14740A, 14CM20312, 14CM20313, 14CM20315 TRIGEN INTERTAN 1.5 11.5MM X 34 CM NAIL, 130o, RIGHT, REF 71676576, Lot Numbers: 14CM19635, 14CM19636, 14DM06203 TRIGEN INTERTAN 1.5 11.5MM X 36 CM NAIL, 130o, LEFT, REF 71676577, Lot Numbers: 14CM17887 TRIGEN INTERTAN 1.5 11.5MM X 36 CM NAIL, 130o, RIGHT, REF 71676578, Lot Numbers: 14CM19637, 14CM19638 TRIGEN INTERTAN 1.5 11.5MM X 38 CM NAIL, 130o, LEFT, REF 71676579, Lot Numbers: 14CM14070, 14CM14071, 14CM14072, 14CM17970, 14DM13200, 14DM13202 TRIGEN INTERTAN 1.5 11.5MM X 38 CM NAIL, 130o, RIGHT, REF 71676580, Lot Numbers: 14CM14073, 14DM14011, 14CM14074 TRIGEN INTERTAN 1.5 11.5MM X 40 CM NAIL, 130o, LEFT, REF 71676581, Lot Numbers: 14CM20367, 14CM20368, 14DM11027 TRIGEN INTERTAN 1.5 11.5MM X 40 CM NAIL, 130o, RIGHT, REF 71676582, Lot Numbers: 14CM14077, 14CM14078, 14CM20369, 14CM20370, 14CM20371, 14CM20372 TRIGEN INTERTAN 1.5 11.5MM X 42 CM NAIL, 130o, LEFT, REF 71676583, Lot Numbers: 13JM17934, 14DM16822 TRIGEN INTERTAN 1.5 11.5MM X 42 CM NAIL, 130o, RIGHT, REF 71676584, Lot Numbers: 14DM06204, 14CM21734, 14CM21733 TRIGEN INTERTAN 1.5 11.5MM X 44 CM NAIL, 130o, LEFT, REF 71676585, Lot Numbers: 14DM07994 TRIGEN INTERTAN 1.5 13MM X 34 CM NAIL, 125o, LEFT, REF 71676597, Lot Numbers: 14DM02923 TRIGEN INTERTAN 1.5 13MM X 34 CM NAIL, 125o, RIGHT, REF 71676598, Lot Numbers: 14DM07995 TRIGEN INTERTAN 1.5 13MM X 38 CM NAIL, 125o, RIGHT, REF 71676602, Lot Numbers: 14DM08001, 14DM11954 TRIGEN INTERTAN 1.5 13MM X 40 CM NAIL, 125o, LEFT, REF 71676603, Lot Numbers: 14DM14038 TRIGEN INTERTAN 1.5 13MM X 42 CM NAIL, 125o, RIGHT, REF 71676606, Lot Numbers: 14CM16792 TRIGEN INTERTAN 1.5 13MM X 44 CM NAIL, 125o, RIGHT, REF 71676608, Lot Numbers: 14CM16793 TRIGEN INTERTAN 1.5 13MM X 32 CM NAIL, 130o, RIGHT, REF 71676618, Lot Numbers: 14CM16796 TRIGEN INTERTAN 1.5 13MM X 34 CM NAIL, 130o, RIGHT, REF 71676620, Lot Numbers: 14DM16271 TRIGEN INTERTAN 1.5 13MM X 36 CM NAIL, 130o, RIGHT, REF 71676622, Lot Numbers: 14CM01174 TRIGEN INTERTAN 10MM X 24 CM NAIL, 125o, REF 71934479, Lot Numbers: 14CM11765 TRIGEN INTERTAN 10MM X 24 CM NAIL, 130o, REF 71934480, Lot Numbers: 14CM11766 TRIGEN INTERTAN 11.5MM X 24 CM NAIL, 130o, REF 71934482, Lot Numbers: 14CM11768

Distribution

Worldwide Distribution - US Nationwide and the countries of AUSTRALIA, CANADA, CHINA, INDIA, UAE, JAPAN, SOUTH AFRICA, SOUTH KOREA, UNITED KINGDOM. AUSTRIA, NORWAY, POLAND, PORTUGAL

Quantity

601 units