FDA Enforcement Class II Ongoing

Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium , Delta Professional , and Delta-Pl us.

Recall: Z-2255-2016 · Reported August 10, 2016

Enforcement

Recall Number
Z-2255-2016
Event ID
74593
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Olympus Scientific Solutions Americas
Voluntary / Mandated
FDA Mandated
Initial Notification
Letter
Report Date
August 10, 2016
Initiation Date
June 22, 2016
Classification Date
July 29, 2016
Address
48 Woerd Ave, Waltham, MA, 02453-3824, United States

Description

Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, Delta, Delta 50, Delta Classic Plus, Delta Element, Delta Premium , Delta Professional , and Delta-Pl us.

Reason

Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.

Code Info

All Models of the Delta XRF Analyzer with Software version 2.5.17.1 and higher

Distribution

Nationwide Distribution.

Quantity

2, 210 units distributed in the USA