Simpleware ScanIP software. For transfer of imaging information from a medical scanner to an output file, and for simulating/evaluating surgical treatment options.
Enforcement
- Recall Number
- Z-2253-2021
- Event ID
- 88207
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Simpleware Product Group, SYNOPSYS NORTHERN EUROPE
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 18, 2021
- Initiation Date
- June 4, 2021
- Classification Date
- August 12, 2021
- Termination Date
- September 29, 2022
- Address
- Bradninch Hall, Castle Street, Exeter, N/A, N/A, United Kingdom
Description
Simpleware ScanIP software. For transfer of imaging information from a medical scanner to an output file, and for simulating/evaluating surgical treatment options.
A issue (bug) has been identified with the interface and image software which could result in anatomical orientation tags/labels being displayed incorrectly. This could result in operator misunderstanding and errors during resampling or re-alignment of image data
Affected version/UDI Simpleware ScanIP 7.0: Created before UDI-DI system (released December 2nd, 2014), Simpleware ScanIP 2016.09: 00863520000307, Simpleware ScanIP M-2017.06: 00863520000314, Simpleware ScanIP N-2018.03: 00863520000321
US: Nationwide Distribution in the states of CA, FL and countries of : France, United Kingdom, Germany, Australia
4 softwares