FDA Enforcement Class II Terminated

Pivot Bipolar Cup 22x42/43-51/53mm, REF 133-YYYY Product Usage: Used in a hemiarthroplasty procedure as the bearing surface to articulate with both the CoCr femoral head and the natural acetabulum

Recall: Z-2253-2014 · Reported August 27, 2014

Enforcement

Recall Number
Z-2253-2014
Event ID
68898
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Ortho Development Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 27, 2014
Initiation Date
July 11, 2014
Classification Date
August 17, 2014
Termination Date
October 6, 2014
Address
12187 Business Park Dr, Draper, UT, 84020-8663, United States

Description

Pivot Bipolar Cup 22x42/43-51/53mm, REF 133-YYYY Product Usage: Used in a hemiarthroplasty procedure as the bearing surface to articulate with both the CoCr femoral head and the natural acetabulum

Reason

Ortho Development is recalling various Pivot Bipolar Cup, Patella, PS Tibial Insert, UC Tibila Insert, and CK Tibil Insert because the foil pouches did not properly seal, allowing air to seep back into the pouch after sealing which increases the risk that the material properties of the implant would be substantially degraded due to oxidation.

Code Info

Pivot Bipolar Cup 22x42mm,Part 133-2242,Lot A136254,Pivot Bipolar Cup 22x43mm,Part 133-2243,Lot A136256,Pivot Bipolar Cup 22x43mm,Part 133-2243,Lot A136255,Pivot Bipolar Cup 22x44mm,Part 133-2244,Lot A136257,Pivot Bipolar Cup 22x45mm,Part 133-2245,Lot A136258,Pivot Bipolar Cup 22x46mm,Part 133-2246,Lot A136259,Pivot Bipolar Cup 22x46mm,Part 133-2246,Lot A136260,Pivot Bipolar Cup 22x47mm,Part 133-2247,Lot A136261,Pivot Bipolar Cup 22x48mm,Part 133-2248,Lot A136262,Pivot Bipolar Cup 22x49mm,Part 133-2249,Lot A136263,Pivot Bipolar Cup 22x50mm,Part 133-2250,Lot A136265,Pivot Bipolar Cup 22x51mm,Part 133-2251,Lot A136266,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136269,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136268,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136267,Pivot Bipolar Cup 22x53mm,Part 133-2253,Lot A136270,

Distribution

Worldwide Distribution - US Nationwide and the country of Japan. No military/govt/VA consignees.

Quantity

188