Philips IntelliSpace PACS 4.4 image management systems performs digital image processing, measurement and communication and storage.
Recall: Z-2252-2014
·
Reported August 27, 2014
Enforcement
- Recall Number
- Z-2252-2014
- Event ID
- 68979
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Healthcare Informatics, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 27, 2014
- Initiation Date
- July 24, 2014
- Classification Date
- August 15, 2014
- Termination Date
- February 11, 2016
- Address
- 4100 E 3rd Ave Ste 101, Foster City, CA, 94404-4819, United States
Description
Philips IntelliSpace PACS 4.4 image management systems performs digital image processing, measurement and communication and storage.
Reason
A problem was detected in the Philips IntelliSpace PAC 4.4 software where if patient information is changed in the DICOM header and the study is exported before the image is updated in the PACS, the unchanged data may export causing the potential for diagnosis on incomplete data.
Code Info
IntelliSpace 4.4
Distribution
US and Australia, Belgium, Canada, Egypt, France, Germany, Netherlands, Oman, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom.
Quantity
109 units