The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. Product Usage: The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories
Enforcement
- Recall Number
- Z-2245-2012
- Event ID
- 62790
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Philips Medical Systems (Cleveland) Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 29, 2012
- Initiation Date
- August 3, 2012
- Classification Date
- August 22, 2012
- Termination Date
- July 23, 2015
- Address
- 595 Miner Road, N/A, Cleveland, OH, 44143-2131, United States
Description
The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. Product Usage: The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories
The patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.
System Code # 728243, Serial #s: 7392, 7394, 7399, 7405, 7406, 7429, 7433, 7440, 7457, 7465, 7469, 7479, 7481, 7487, 7488, 7491, 7502, 7506, 7507, 7517, 7518, 7520, 7521, 7532, 7539, 7545, 7547, and 7556. System Code # 728244, Serial #s: 7190, 7195, 7318, 7383, 7427, 7430, 7510, and 7533. System Code #728256, Serial #s: 3163, 3500, 3500, 5623, 5655, 5655, 5666, 5670, 5715, 5724, 5747, 5773, 5776, 5813, 5836, 5855, 5862, 5868, 5877, 5897, 5911, 5911, 5916, 5935, 5969, and 5997. System Code #728251, Serial #: 4051. System Code # 728246, Serial #s: 5547, 5585, 5791, 5803, 5876, 5974. System Code # 728235, Serial #s: 9233, 9234, 9261, 95022. System Code # 728231, Serial #s: 9867, 9925, 9931, 9980, 9996, 10009, 10011, 10015, 10028, 10036, 10042, 10045, 10050, 10052, 10062, 10063, 10064, 10077, 10079, 10081, 10096, 10099, 10102, 10111, 10114, 10120, 10122, 10133, 10135, 10136, 10137, 10139, 10145, 10148, 10149, 10151, 10153, 10166, 10170, 10172, 10198, 10203, 95109, 95348, 95563.
Worldwide Distribution - US Nationwide including the states of: AL, AR, AZ, CA, DC, FL, GA, KY, MD, MI, MN, MS, ND, NM, NY, OH, OK, SC, TN, TX, VT, WA, and WI and the countries of: Austria, China, Denmark, France, Germany, Iceland, Japan, Netherlands, South Africa, Sweden, Switzerland, United Kingdom. There were two government consignees. There were no Canadian consignees.
126 units