FDA Enforcement Class I Terminated

Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201

Recall: Z-2244-2021 · Reported August 25, 2021

Enforcement

Recall Number
Z-2244-2021
Event ID
88266
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Cardinal Health 200, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 25, 2021
Initiation Date
June 16, 2021
Classification Date
August 16, 2021
Termination Date
January 4, 2023
Address
15 Hampshire St Bldg 5, N/A, Mansfield, MA, 02048-1113, United States

Description

Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201

Reason

The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.

Code Info

All lot codes

Distribution

US Nationwide distribution.

Quantity

7197 units