Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin Antibody II, REF A32898 (UDI: 15099590202910) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.
Enforcement
- Recall Number
- Z-2242-2020
- Event ID
- 84862
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 10, 2020
- Initiation Date
- April 30, 2019
- Classification Date
- June 2, 2020
- Termination Date
- December 28, 2021
- Address
- 250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States
Description
Beckman Coulter ACCESS Immunoassay Systems, Access Thyroglobulin Antibody II, REF A32898 (UDI: 15099590202910) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.
Catalog Number: A32898; all lots.
US Distribution: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusett, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico New York, North Carolina, North Dakota, Ohio, Oklahoma Oregon, Pennsylvania, Rhode Island, South Carolina,South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, and Wyoming. US Territories: American Samoa, Guam, and Puerto Rico. OUS Distribution: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Djibouti, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Liberia, Lithuania, Malawi, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Antillean, New Zealand, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, San Marino, Serbia, Sierra Leone, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Zambia, and Zimbabwe.
123,324 units