FDA Enforcement Class II Terminated

Dimension Vista¿ Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative measurement of total homocysteine in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System.

Recall: Z-2241-2016 · Reported July 27, 2016

Enforcement

Recall Number
Z-2241-2016
Event ID
74468
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
CSL Behring GmbH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 27, 2016
Initiation Date
May 17, 2016
Classification Date
July 19, 2016
Termination Date
February 7, 2018
Address
Emil-von-Behring-Str. 76, N/A, Marburg, N/A, N/A, Germany

Description

Dimension Vista¿ Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative measurement of total homocysteine in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System.

Reason

Reagent lots # 15243MA and 15243MB shows a reduced stability once opened that does not meet the Instructions For Use claim of seven (7) days. A maximum bias of -37% was observed.

Code Info

Lot Number 15243MA Mfg Date: 08/31/2015 Exp Date: 05/31/2016 Lot Number 15243MB Mfg Datge: 08/31/2015 Exp Date: 05/31/2016

Distribution

Distributed to: AL, AR, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WI, WV

Quantity

898 units