GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused protection of internal system components from the high voltage system components, such as the X-ray tube or High Voltage Transformer. System model number A349855, Snubber Board number 00-880405-03 rev 5. This product is used for image intensified fluoroscopic x-rays for interventional radiologic procedures.
Enforcement
- Recall Number
- Z-2241-2012
- Event ID
- 62729
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- GE OEC Medical Systems, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Visit
- Report Date
- August 29, 2012
- Initiation Date
- April 26, 2010
- Classification Date
- August 22, 2012
- Termination Date
- August 22, 2012
- Address
- 384 Wright Brothers Dr, N/A, Salt Lake City, UT, 84116-2862, United States
Description
GE HC OEC 9900 Elite and GE OEC 9800 Fluoroscopes Snubber Boards are a PCB with a 70 amp fuse to provide fused protection of internal system components from the high voltage system components, such as the X-ray tube or High Voltage Transformer. System model number A349855, Snubber Board number 00-880405-03 rev 5. This product is used for image intensified fluoroscopic x-rays for interventional radiologic procedures.
GE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
Serial numbers: 69-0128, 82-0157, 82-0158, 82-0185, 82-0191, 82-0210, 82-0215, 82-0223, 82-0229, 82-0266, 82-0324, 82-0328, 82-0464, 82-0477, 82-0484, 82-0505-RC, 82-0507, 82-0542, 82-0544, 82-0546, 82-0551, 82-0604, 82-0648, 82-0673, 82-0680, 82-0714, 82-0733, 82-0779, 82-0858, 82-0869, 82-0884, 82-0906, 82-0918, 82-0965, 82-1004, 82-1009, 82-1057-RC, 82-1070, 82-1139, 82-1191, 82-1194, 82-1225, 82-1247, 82-1303, 82-1405, 82-1434, 82-1458, 82-1480, 82-1534, 82-1547-RC, 82-1564, 82-1585, 82-1588, 82-1601, 82-1604, 82-1621, 82-1694, 82-1695, 82-1699, 82-1797, 82-1798, 82-1842, 82-1850, 82-1921, 82-1959, 82-2015, 82-2120, 82-2192, 82-2245, 82-2247, 82-2274, 82-2277, 82-2306, 82-2386, 82-2487, 82-2661, 82-2662, 82-2716, 82-2721, 82-2741, 82-2745, 82-2797, 82-2846, 82-2905, 82-2965, 82-3102, 82-3107, 82-3122, 82-3135, 82-3151, 82-3351, 82-3366, 82-3403, 82-3431, 82-3492, 82-3498, 82-3563, 82-3570, 82-3614, 82-3634, 82-3762, 82-3815, 82-3819, 82-3824, 82-3839, 82-3851, 82-3882, 82-3884, 82-3887, 82-3888, 82-3890, 82-3894, 82-3895, 82-3896, 82-3898, 82-3899, 82-3901, 82-3903, 82-3904, 82-3910, 82-3911, 82-3912, 82-3913, 82-3914, 82-3915, 82-3917, 82-3919, 82-3920, 82-3921, 82-3922, 82-3923, 82-3925, 82-3928, 82-3929, 82-3930, 82-3931, 82-3933, 82-3934, 82-3935, 82-3936, 82-3937, 82-3938, 82-3939, 82-3941, 82-3942, 82-3943, 82-3944, 82-7011-MH, 82-7024-MH, 82-7154-CMH, 82-7189-MH, 82-7190-MH, 82-7191-MH, 82-7197-MH, 82-7200-MH, 82-7206-MH, 89-0100, 89-0128, 89-0135-RC, 89-0210, 89-0248, 89-0315, 89-0362, 89-0385, 89-0401, 89-0416, 89-0457, 89-0464, 89-0467, 89-0505, 89-0506, 89-0537, 89-0556, 89-0559, 89-0577, 89-0612, 89-0643, 89-0660, 89-0662, 89-0729, 89-0738, 89-0780, 89-0781, 89-0832, 89-0853, 89-0868, 89-0876, 89-0882, 89-0922, 89-0954-RC, 89-0963, 89-1040, 89-1071, 89-1172, 89-1195, 89-1228, 89-1245, 89-1248, 89-1255, 89-1262, 89-1268, 89-1271, 89-1274, 89-1286, 89-1295, 89-1308, 89-1343, 89-1373, 89-1400-RC, 89-1408, 89-1426, 89-1494, 89-1531, 89-1576, 89-1598, 89-1606, 89-1612, 89-1640, 89-1663, 89-1673, 89-1721, 89-1729, 89-1739, 89-1749, 89-1868, 89-2121, 89-2259, 89-2264, 89-2275, 89-2309, 89-2315, 89-2350, 89-2354, 89-2456, 89-2464, 89-2500, 89-2680, 89-2866, 89-3211, 89-3294, 89-3312, 89-3349, 89-3352, 89-3355, 89-3377, 89-3420, 89-3433, 89-3461, 89-3479, 89-3483, 89-3488, 89-3490, 89-3494, 89-3495, 89-3499, 89-3501, 89-3504, 89-3508, 89-3509, 89-3511, 89-3512, 89-3513, 89-3514, 89-3515, 89-3519, 89-3520, 89-3522, 89-3524, 89-3527, 89-3529, 89-3530, 89-3533, 89-3537, 89-3538, 89-3539, 89-3543, 89-3544, 89-3545, 89-3546, 89-3547, 89-3548, 89-3549, 89-3551, 89-3552, 89-3553, 89-3554, 89-3555, 89-3556, 89-3557, 89-3558, 89-3559, 89-3560, 89-3561, 89-3562, 89-3563, 89-3564, 89-3565, 89-3567, 89-3568, 89-3569, 89-3570, 89-3571, 89-3572, 89-3573, 89-3574, 89-3578, 89-3579, 8S-0027, 8S-0076-RC, 8S-0086, 8S-0121-C, 8S-0158, 8S-0165, 8S-0245, 8S-0291, 8S-0301-C, 8S-0315, 8S-0413-C, 8S-0494, 8S-0615-C, 8S-0642, 8S-0691, 8S-0707, 8S-0711, 8S-0735, 8S-0747, 8S-0776, 8S-0840, 8S-0844-RC, 8S-0858, 8S-0867, 8S-0872, 8S-0921, 8S-0974, 8S-1013-C, 8S-1046, 8S-1056, 8S-1058, 8S-1065-C, 8S-1174, 8S-1203, 8S-1224-C-RC, 8S-1285-RC, 8S-1360, 8S-1445, 8S-1644, 8S-1856, 8S-1930-N, 8S-1954, 8S-1959, 8S-1961, 8S-2052, 8S-2053, 8S-2063, 8S-2063, 8S-2113, 8S-2145, 8S-2164, 8S-2170, 8S-2207, 8S-2256, 8S-2329, 8S-2439, 8S-2451, 8S-2452, 8S-2452, 8S-2462, 8S-2532, 8S-2622, 8S-2865, 8S-2956, 8S-2985, 8S-3002, 8S-3108, 8S-3111, 8S-3112, 8S-3130, 8S-3179, 8S-3183, 8S-3197, 8S-3205, 8S-3207, 8S-3211-CH, 8S-3216, 8S-3222, 8s-3231, 8S-3233, 8S-3243, 8S-3245, 8S-3246, 8S-3247, 8S-3248, 8S-3249, 8S-3253, 8S-3254, 8S-3255, 8S-3256-H, 8S-3257, 8S-3258-CH, 8S-3259, 8S-3260, 8S-3261, 8S-3262, 8S-3263, 8S-3264, 8S-3265, 8S-3266, 8S-3267, 8S-3268, 8S-3269-H, 8S-3270-H, 8S-3271-H, 8S-3272-H, 8S-3273-CH, 8S-3274-CH, 8S-3275, 8S-3276, 8S-3278, 8S-3279, 8S-3280, 8S-3281, 8S-3282, 8S-3283, 8S-3285, 8S-3286, 8S-3287, 8S-3288-H, 8S-3289-H, 8S-3291, 8S-3292, 8S-3296, 8S-3299, 8S-3301-H, 8S-3302, 8S-3305, 8S-3306, 8S-3307, 8S-3308, 8S-3309, 8S-3310-H, 8S-3311, 8S-3315, 8S-7094-CMH, 8S-7228-CMH, 8S-7229-CMH, 8S-7230-CMH, 8S-7231-MH, 8S-7232-MH, 8S-7233-MH, E2-0001, E2-0015, E2-0026, E2-0028, E2-0034, E2-0078, E2-0084, E2-0085, E2-0108, E2-0110, E2-0111, E2-0117, E2-0120, E2-0122, E2-0130, E2-0142, E2-0143, E2-0148, E2-0151, E2-0152, E2-0153, E2-0154, E2-0155, E2-0156, E2-0157, E2-0158, E2-0159, E2-0160, E2-0161, E2-0162, E2-0163, E2-0164, E2-0165, E2-0167, E2-0168, E2-0169, E2-0171, E2-0172, E2-0173, E2-0174, E2-0175, E2-0176, E2-0177, E2-0178, E2-0179, E2-0180, E2-0181, E2-0182, E2-0183, E2-0184, E2-0185, E2-0186, E2-0187, E2-0188, E2-0189, E2-0190, E2-0191, E2-0192, E2-0193, E2-0194, E2-0195, E2-0196, E2-0197, E2-0198, E2-0199, E2-0200, E2-0201, E2-0202, E2-0203, E2-0204, E2-0205, E2-0206, E2-0207, E2-0208, E2-0209, E2-0210, E2-0211, E2-0212, E2-0213, E2-0215, E2-0216, E2-0217, E2-0222, E2-0224, E2-0225, E2-0230, E2-0231, E2-0232, E2-0233, E2-0234, E2-0235-C, E2-0237, E2-0238, E2-0239, E2-0240, E2-0241, E2-0242, E2-0243, E2-0244, E2-0245, E2-0246, E2-0247, E2-0248, E2-0250, E2-0251, E2-0252, E2-0253, E2-0254, E2-0255, E2-0256, E2-0257, E2-0258, E2-0259, E2-0260, E2-0261, E2-0262, E2-0263, E2-0264, E2-0265, E2-0266, E2-0267, E2-0268, E2-0269, E2-0270, E2-0271, E2-0272, E2-0273, E2-0274, E2-0275, E2-7014-MH, E2-7020-MH, E2-7048-MH, E2-7050-MH, E2-7051-MH, E2-7057-MH, E2-7058-MH, E2-7059-MH, E2-7060-MH, E2-7061-MH, E2-7062-MH, E2-7063-MH, E2-7064-MH, E2-7065-MH, E2-7066-MH, E2-7067-MH, E2-7068-MH, E2-7069-CMH, E2-7070-MH, E2-7071-MH, E2-7072-MH, E2-7074-MH, E2-7075-MH, E2-7076-MH, E2-7077-CMH, E2-7078-MH, E2-7079-MH, E9-0008, E9-0025, E9-0031, E9-0035-G, E9-0037, E9-0050, E9-0083, E9-0085, E9-0086, E9-0088, E9-0089-G, E9-0090, E9-0091, E9-0092, E9-0093, E9-0094, E9-0095, E9-0098, E9-0099, E9-0100, E9-0101, E9-0102, E9-0103, E9-0108, E9-0109, E9-0110, E9-0111, E9-0112, E9-0113, E9-0114, E9-0115, E9-0116, E9-0118, E9-0119, E9-0120, E9-0121, E9-0122, E9-0123, E9-0124, E9-0125, E9-0126, E9-0127, E9-0128, E9-0129, E9-0129, E9-0130, E9-0139, E9-0141, E9-0142, E9-0143, E9-0144, E9-0145, E9-0146, E9-0147, E9-0148, E9-0149, E9-0150, E9-0151, E9-0152, E9-0153, E9-0155, E9-0157, E9-0158, E9-0159, E9-0162, E9-0163, E9-0164, E9-0165, E9-0166, E9-0169, E9-0170, E9-0171, E9-0172, E9-0173, E9-0177, E9-0178, E9-0179, E9-0180, E9-0181, E9-0183, E9-0184, E9-0185, E9-0188, E9-0249, E9-0253, ES-0005, ES-0006, ES-0007-C, ES-0022-CH, ES-0029-CH, ES-0034-G, ES-0056, ES-0069, ES-0071, ES-0072, ES-0076, ES-0078, ES-0084-G, ES-0086-G, ES-0087, ES-0088, ES-0089, ES-0090, ES-0091-H, ES-0092-CH, ES-0096, ES-0097, ES-0098, ES-0099-CH, ES-0101-CH, ES-0102-CH, ES-0103, ES-0104, ES-0105, ES-0106, ES-0107, ES-0108, ES-0109, ES-0110, ES-0111, ES-0112, ES-0113, ES-0114, ES-0115, ES-0116, ES-0117, ES-0118, ES-0119, ES-0120, ES-0121, ES-0122, ES-0123, ES-0124, ES-0125, ES-0126, ES-0127, ES-0128, ES-0129, ES-0130, ES-0131-H, ES-0132, ES-0133, ES-0134, ES-0135, ES-0136, ES-0138, ES-0139, ES-0140, ES-0141, ES-0142, ES-0147, ES-0152, ES-0155-H, ES-0156, ES-0158, ES-0159, ES-0160, ES-0163-G, ES-0169-CH, ES-0170, ES-0171, ES-0172, ES-0176, ES-0178, ES-0180, ES-0181, ES-0182, ES-0184-CH, ES-0185-H, ES-0186-H, ES-0187-H, ES-0190-CH, ES-0191-H, ES-0192-H, ES-7010-CMH, ES-7015-MH, ES-7018-MH, ES-7021-MH, ES-7028-CMH, ES-7029-MH, ES-7030-CMH, ES-7031-CMH, ES-7033-CMH, ES-7034-CMH, ES-7035-CMH, ES-7036-CMH, ES-7037-CMH, ES-7039-MH.
Worldwide and Nationwide (USA) distribution.
845 units