Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088
Enforcement
- Recall Number
- Z-2240-2023
- Event ID
- 92622
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 2, 2023
- Initiation Date
- June 5, 2023
- Classification Date
- July 27, 2023
- Address
- 40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States
Description
Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088
Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of liquid following the plexiglass dropping off, at the same time the user or operator happens to touch the liquid on the cover which is connected to the internal live parts).
UDI-DI :04056869148243 US S/N: 30001 30002 30008 30013 30029 30031 30032 30033 30034
Worldwide - US Nationwide distribution in the states of FL, GA, KS, MN, NC, NY and the countries of Austria, Bangladesh, Brazil, Canada, China, Croatia, El Salvador, France, Germany, Indonesia, Italy, Japan, Jordan, Lebanon, Luxembourg, Mauritius, Mexico, Monaco, Peru, Philippines, Poland, Romania, Slovakia, South Africa, Spain, Switzerland, Taiwan, Ukraine, United Kingdom.
9 units