3.2MM X 510MM COCR THD TIP; 14-443053 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
Enforcement
- Recall Number
- Z-2240-2018
- Event ID
- 80193
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- June 27, 2018
- Initiation Date
- May 16, 2018
- Classification Date
- June 15, 2018
- Termination Date
- April 5, 2024
- Address
- 56 E Bell Dr, N/A, Warsaw, IN, 46582-6989, United States
Description
3.2MM X 510MM COCR THD TIP; 14-443053 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.
001390 105790 324290 974210 820900 512140 428260 915580 017930 671970 055480 615280 506500 569970 053580 367190 932620 015010 141060 820240 941850 682590 401830
Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.