FDA Enforcement Class II Ongoing

BD Trucount Controls, 30 Test - REF: 340335

Recall: Z-2238-2025 · Reported August 13, 2025

Enforcement

Recall Number
Z-2238-2025
Event ID
97141
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Becton, Dickinson and Company, BD Bio Sciences
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 13, 2025
Initiation Date
July 8, 2025
Classification Date
August 1, 2025
Address
155 N Mccarthy Blvd, N/A, Milpitas, CA, 95035-5102, United States

Description

BD Trucount Controls, 30 Test - REF: 340335

Reason

Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.

Code Info

Lot: 85708 UDI: 00382903403356 *Distributed outside the US

Distribution

Worldwide - US Nationwide distribution in the states of Tennessee, New York, Massachusetts and the countries of Argentina, Belgium, Botswana, Brazil, Canada, France, Germany, Lithuania, Peru, Philippines, United Kingdom.

Quantity

629 kits