Ringed DxI Reaction Vessels (RVs)
Enforcement
- Recall Number
- Z-2237-2023
- Event ID
- 92576
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Beckman Coulter, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 2, 2023
- Initiation Date
- June 20, 2023
- Classification Date
- July 26, 2023
- Address
- 1000 Lake Hazeltine Dr, N/A, Chaska, MN, 55318-1037, United States
Description
Ringed DxI Reaction Vessels (RVs)
Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
UDI: 15099590229979; Catalog Number: 386167; Lot Numbers: L17855111, L17798598, L18046334
Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.
213 cases of 10,000 RVs (2,130,000 RVs)