FDA Enforcement Class II Ongoing

Ringed DxI Reaction Vessels (RVs)

Recall: Z-2237-2023 · Reported August 2, 2023

Enforcement

Recall Number
Z-2237-2023
Event ID
92576
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 2, 2023
Initiation Date
June 20, 2023
Classification Date
July 26, 2023
Address
1000 Lake Hazeltine Dr, N/A, Chaska, MN, 55318-1037, United States

Description

Ringed DxI Reaction Vessels (RVs)

Reason

Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.

Code Info

UDI: 15099590229979; Catalog Number: 386167; Lot Numbers: L17855111, L17798598, L18046334

Distribution

Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.

Quantity

213 cases of 10,000 RVs (2,130,000 RVs)