GUIDE WIRE w/ BALL TIP 2.0MMX98CM; 469040 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
Enforcement
- Recall Number
- Z-2234-2018
- Event ID
- 80193
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zimmer Biomet, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- June 27, 2018
- Initiation Date
- May 16, 2018
- Classification Date
- June 15, 2018
- Termination Date
- April 5, 2024
- Address
- 56 E Bell Dr, N/A, Warsaw, IN, 46582-6989, United States
Description
GUIDE WIRE w/ BALL TIP 2.0MMX98CM; 469040 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.
228829 311230 311260 255105 152590 757210 982070 282057 283729 057650 292980 293094 299420 668040 579470 354970 519980 670820 695640 963100
Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.