RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
Enforcement
- Recall Number
- Z-2230-2025
- Event ID
- 97209
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Siemens Healthcare Diagnostics Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 6, 2025
- Initiation Date
- June 26, 2025
- Classification Date
- July 31, 2025
- Address
- 2 Edgewater Dr, Norwood, MA, 02062-4637, United States
Description
RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.
worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
1,477 units