FDA Enforcement Class II Ongoing

RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

Recall: Z-2230-2025 · Reported August 6, 2025

Enforcement

Recall Number
Z-2230-2025
Event ID
97209
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Siemens Healthcare Diagnostics Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 6, 2025
Initiation Date
June 26, 2025
Classification Date
July 31, 2025
Address
2 Edgewater Dr, Norwood, MA, 02062-4637, United States

Description

RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

Reason

Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.

Code Info

Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.

Distribution

worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.

Quantity

1,477 units