Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
Enforcement
- Recall Number
- Z-2228-2017
- Event ID
- 77282
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- ConvaTec, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 14, 2017
- Initiation Date
- May 10, 2017
- Classification Date
- June 2, 2017
- Termination Date
- August 27, 2021
- Address
- 7900 Triad Center Dr Ste 400, N/A, Greensboro, NC, 27409-9076, United States
Description
Oxygen Masks, Elongated (under chin), Adult Med Conc., No Tubing, Product Code 104-E;
Reports of tubing detaching from oxygen masks either prior to use (in packaging) or during use were made by different end users and for different batches. The failures were caused by an improper solvent bond between the oxygen tubing and the oxygen mask connector. While the defect affects less than 2% of recalled product, there is a potential for harm when oxygen supply to the user is disrupted.
Lot Number 102153, 102339, 102941, 103176, 103453, 103718, 103919, 104471, 105034, 105282, 105427, 105567, 105996, 106960, 108658, 112024, 112530
CA, CT, FL, GA, IL, MO, NJ, NY, NC, OH, PA, RI, VA , Puerto Rico, Australia, Belgium, Canada, Columbia, Czech, Denmark, Finland, France, Germany, Greece, Israel, Italy, Japan, Jordan, Malaysia, Malta, Netherlands, Norway, Rwanda, Saudi Arabia, Sweden, Switzerland, United Kingdom, and Vietnam
816,709